A Hand-Guided Robotic Drill for Vestibular Implant Surgery—Feasibility of Preventing Membranous Labyrinth Rupture
Open Access
- 8 May 2026
- journal article
- research article
- Published by SAGE Publications in Journal of Otolaryngology - Head & Neck Surgery
- Vol. 55
Abstract
Importance: Progress in vestibular implantation offers hope for patients with bilateral vestibulopathy. However, surgically opening the semicircular canals risks breaching the membranous labyrinth, which may induce sensorineural hearing loss. A robotic drill sensing force and torque might prevent membranous labyrinth rupture. Primary objective: To assess the feasibility of force- and torque-based automatic cessation in a hand-guided robotic drill for fenestrating the bony semicircular canals without rupturing the membranous labyrinth. Secondary objective: To fit an electrode dummy through the fenestrations. Design: Feasibility study using human cadaveric temporal bones. Setting: Laboratory. Participants: Ten formalin-fixed human temporal bones. Intervention: After performing a cortical mastoidectomy and skeletonizing the semicircular canals, a hand-guided robotic drill was used to drill 2 fenestrations in each semicircular canal. A silicone electrode dummy was inserted through each fenestration. Main outcome measures: Proportion of fenestrations with intact membranous labyrinth, as evaluated with a surgical microscope. Proportion of fenestrations allowing electrode insertion without additional manipulation. Results: A total of 60 fenestrations were made in 30 semicircular canals from 10 temporal bones. Technical issues related to drill bit fixation occurred in 6 fenestrations. The remaining 54 fenestrations were all made without visible damage to the membranous labyrinth. In 81% of these fenestrations (44/54), the electrode could be advanced without requiring additional manipulation. The technical issue was related to improper alignment of the drill bit, leading to incorrect force and torque sensing. Conclusions: Force- and torque-based automatic cessation in a hand-guided robotic drill is feasible for fenestrating the bony semicircular canals without rupturing the membranous labyrinth. However, improved burr fixation is required for consistent and reliable performance. Relevance: The investigated approach holds potential to improve safety and precision in semicircular canal surgery, such as vestibular implantation. This may expand treatment options for patients with residual inner ear function.Keywords
This publication has 30 references indexed in Scilit:
- A first exploration of the economic consequences of an autonomous surgical robot for lateral skull base surgery: an early health technology assessmentInternational Journal of Technology Assessment in Health Care, 2023
- The Next Challenges of Vestibular Implantation in HumansJournal of the Association for Research in Otolaryngology, 2023
- The resilience of the inner ear—vestibular and audiometric impact of transmastoid semicircular canal pluggingJournal of Neurology, 2021
- “Sandwich technique” enables preservation of hearing and antivertiginous effect in cholesteatomatous labyrinthine fistulaEuropean Archives of Oto-Rhino-Laryngology, 2021
- A case of perilymphatic fistula caused by surgical drilling with preserved normal hearingAmerican Journal of Otolaryngology, 2021
- Posture, Gait, Quality of Life, and Hearing with a Vestibular ImplantNew England Journal of Medicine, 2021
- Robotic Ear SurgeryOtolaryngologic Clinics of North America, 2020
- Vestibular Implantation and the Feasibility of Fluoroscopy-Guided Electrode InsertionOtolaryngologic Clinics of North America, 2020
- Bilateral Vestibular DeficiencyJAMA Otolaryngology–Head & Neck Surgery, 2014
- Prevalence and Impact of Bilateral Vestibular HypofunctionJAMA Otolaryngology–Head & Neck Surgery, 2013