Agreement between Spirometers: A Challenge in the Follow-Up of Patients and Populations?
Authors:
- M.W. GerbaseMGM.W. Gerbase
- E. Dupuis-LozeronEDE. Dupuis-Lozeron
- C. SchindlerCSC. Schindler
- D. KeidelDKD. Keidel
- P.O. BridevauxPBP.O. Bridevaux
- S. KriemlerSKS. Kriemler
- N.M. Probst-HenschNPN.M. Probst-Hensch
- T. RochatTRT. Rochat
- N. KünzliNKN. Künzli
Open Access
- 8 March 2013
- journal article
- research article
- Published by S. Karger AG in Respiration
- Vol. 85 (6) , 505-514
Abstract
Background: Long-term cohort studies and lung function laboratories are confronted with the need for replacement of spirometers. Lack of agreement between spirometers might affect the longitudinal comparison of data, notably when replacing conventional by portable spirometers. Objectives: To compare the handheld EasyOne (EO) with the conventional SensorMedics (SM) spirometer, and to analyze the interdevice reproducibility of EO spirometers. Methods: In total, 82 volunteers completed spirometry sessions with 1 SM and 2 of 3 EO spirometers following a Latin square design. Analyses of differences in forced vital capacity (FVC), forced expiratory flow in 1 s (FEV1), FEV1/FVC and mean forced expiratory flow calculated between 25 and 75% of the FVC between spirometers used a mixed effect model with a random intercept for each subject and the effect of the device as fixed effect adjusted for sex, age, height and order of spirometer tested. Bland-Altman plots show the 95% limits of agreement. Results: Comparisons between EO and SM showed relatively small mean differences of 1 in all EO devices. The 95% agreement exceeded the limits for FEV1 by 50 ml in 2 EO spirometers. The EO interdevice comparisons showed mean differences and limits of agreement within established thresholds, thus indicating fair accuracy when comparing devices. Repeats with the same spirometer did not result in statistically significant differences. Conclusions: This study suggests fair agreement between the handheld and the conventional spirometer. Differences slightly exceeding limits for FEV1 in 2 EO devices might be considered mostly irrelevant for clinical practice. However, the systematically lower FVC and FEV1 observed with EO may be significant for epidemiological studies, thus justifying inspection before replacing devices.Keywords
This publication has 14 references indexed in Scilit:
- Accuracy and Precision of Desktop Spirometers in General PracticesRespiration, 2012
- Quality of Spirometry tests performed by 9893 adults in 14 countries: The BOLD StudyRespiratory Medicine, 2011
- Stability of the EasyOne ultrasonic spirometer for use in general practiceRespirology, 2006
- Sources of variation in forced expiratory volume in one second and forced vital capacityEuropean Respiratory Journal, 2006
- Standardisation of spirometryEuropean Respiratory Journal, 2005
- A comparison of spirometry in general practice and a pulmonary function laboratoryPrimary Care Respiratory Journal, 2005
- Hand‐held turbine spirometer: Agreement with the conventional spirometer at baseline and after exercisePediatric Allergy and Immunology, 2005
- SAPALDIA: Methods and participation in the cross-sectional part of the Swiss Study on Air Pollution and Lung Diseases in AdultsInternational Journal of Public Health, 1997
- Lung function in healthy never smoking adults: reference values and lower limits of normal of a Swiss population.Thorax, 1996
- Variability of FVC and FEV1 due to technician, team, device and subject in an eight centre study: three quality control studies in SAPALDIA. Swiss Study on Air Pollution and Lung Disease in AdultsEuropean Respiratory Journal, 1995