Abstract
Objectives:To review the radiation doses received by women attending the UK breast screening programme between 2010 and 2012. To compare doses with previous years and to quantify the impact on dose of changing from analogue to digital imaging and to analyse doses by type of imaging system.Methods:Measurements of doses to samples of about 50-100 women attending for screening were collected across the whole of the UK breast screening programme.Results:Data were collected for 87,122 exposures, using 449 X-ray sets, for 25,408 women. The average MGD was 1.79 mGy for medio-lateral oblique images and 1.58 mGy for cranio-caudal images. The average MGD per two view examination was 4.01 mGy for film-screen imaging and 3.03 mGy for direct digital radiography (DR) and 4.69 for computed radiography (CR).Conclusions:The MGD to women attending breast screening has been reduced on average by about 25% where DR systems have replaced film-screen systems. The dose reduction was greatest for breasts with the largest compres...