Does new magnetic resonance imaging technology provide better geometrical accuracy during stereotactic imaging?
Authors:
- Josef NovotnyJNJosef Novotny
- Josef VymazalJVJosef Vymazal
- Daniela TlachacovaDTDaniela Tlachacova
- Michal SchmittMSMichal Schmitt
- Pavel ChudaPCPavel Chuda
- Dusan UrgosikDUDusan Urgosik
- Roman LiscakRLRoman Liscak
- 1 January 2005
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 102 (Special_Su) , 8-13
Abstract
Object. The authors sought to compare the accuracy of stereotactic target imaging using the Siemens 1T EXPERT and 1.5T SYMPHONY magnetic resonance (MR) units. Methods. A water-filled cylindrical Perspex phantom with axial and coronal inserts containing grids of glass rods was fixed in the Leksell stereotactic frame and subjected to MR imaging in Siemens 1T EXPERT and Siemens 1.5T SYMPHONY units. Identical sequences were used for each unit. The images were transferred to the GammaPlan treatment planning system. Deviations between stereotactic coordinates based on MR images and estimated real geometrical positions given by the construction of the phantom insert were evaluated for each study. The deviations were further investigated as a function of the MR unit used, MR sequence, the image orientation, and the spatial position of measured points in the investigated volume. Conclusions. Larger distortions were observed when using the SYMPHONY 1.5T unit than those with the EXPERT 1T unit. Typical average distortion in EXPERT 1T was not more than 0.6 mm and 0.9 mm for axial and coronal images, respectively. Typical mean distortion for SYMPHONY 1.5T was not more than 1 mm and 1.3 mm for axial and coronal images, respectively. The image sequence affected the distortions in both units. Coronal T2-weighted spin-echo images performed in subthalamic imaging produced the largest distortions of 2.6 mm and 3 mm in the EXPERT 1T and SYMPHONY 1.5T, respectively. Larger distortions were observed in coronal slices than in axial slices in both units, and this effect was more pronounced in SYMPHONY 1.5T. Noncentrally located slice positions in the investigated volume of the phantom were associated with larger distortions. Object. The authors sought to compare the accuracy of stereotactic target imaging using the Siemens 1T EXPERT and 1.5T SYMPHONY magnetic resonance (MR) units. Methods. A water-filled cylindrical Perspex phantom with axial and coronal inserts containing grids of glass rods was fixed in the Leksell stereotactic frame and subjected to MR imaging in Siemens 1T EXPERT and Siemens 1.5T SYMPHONY units. Identical sequences were used for each unit. The images were transferred to the GammaPlan treatment planning system. Deviations between stereotactic coordinates based on MR images and estimated real geometrical positions given by the construction of the phantom insert were evaluated for each study. The deviations were further investigated as a function of the MR unit used, MR sequence, the image orientation, and the spatial position of measured points in the investigated volume. Conclusions. Larger distortions were observed when using the SYMPHONY 1.5T unit than those with the EXPERT 1T unit. Typical average distortion in EXPERT 1T was not more than 0.6 mm and 0.9 mm for axial and coronal images, respectively. Typical mean distortion for SYMPHONY 1.5T was not more than 1 mm and 1.3 mm for axial and coronal images, respectively. The image sequence affected the distortions in both units. Coronal T2-weighted spin-echo images performed in subthalamic imaging produced the largest distortions of 2.6 mm and 3 mm in the EXPERT 1T and SYMPHONY 1.5T, respectively. Larger distortions were observed in coronal slices than in axial slices in both units, and this effect was more pronounced in SYMPHONY 1.5T. Noncentrally located slice positions in the investigated volume of the phantom were associated with larger distortions.Keywords
This publication has 24 references indexed in Scilit:
- An image fusion study of the geometric accuracy of magnetic resonance imaging with the Leksell stereotactic localization systemJournal of Applied Clinical Medical Physics, 2001
- Development of a Unique Phantom to Assess the Geometric Accuracy of Magnetic Resonance Imaging for Stereotactic LocalizationNeurosurgery, 1999
- A three‐dimensional volumetric test object for geometry evaluation in magnetic resonance imagingMedical Physics, 1997
- A Phantom Study to Assess the Accuracy of Stereotactic Localization, Using T1-weighted Magnetic Resonance Imaging with the Leksell Stereotactic SystemNeurosurgery, 1996
- Characterization of Spatial Distortion in Magnetic Resonance Imaging and Its Implications for Stereotactic SurgeryNeurosurgery, 1994
- Quantifying MRI geometric distortion in tissueMagnetic Resonance in Medicine, 1994
- Variables affecting the accuracy of stereotactic localization using computerized tomographyJournal of Neurosurgery, 1993
- A Comparison between Magnetic Resonance Imaging and Computed Tomography for Stereotactic Coordinate DeterminationNeurosurgery, 1992
- Landmark based correlation of medical imagesPhysics in Medicine & Biology, 1990
- Correction of Spatial Distortion in MR ImagingJournal of Computer Assisted Tomography, 1987