Biomechanical performance of the cemented acetabular cup with combined effects of bone quality, implant material combinations and bodyweight
- 22 July 2022
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 236 (9) , 1309-1327
Abstract
The objective of this study is to understand the combined effects of bone quality, implant materials and bodyweight on the biomechanical performance of cemented acetabular cup. Additionally, the performance of the cemented acetabular cup was evaluated for obesity cases or obese people. A total of 84 FE models (based on CT data) were developed based on combinations of three different cancellous bone material distributions to represent bone quality, four different implant material combinations and seven different bodyweights. The biomechanical performance of the acetabular cup was evaluated based on bone stress (both cortical and cancellous bone), cement mantle stress, micromotion and contact pressure between the acetabular cup and femoral head. Cortical bone stress, cancellous bone stress, cement stress, the contact pressure between implants and micromotion between implants are affected by different bone quality, implant material combinations and bodyweights. An increase in bodyweight would increase the cortical bone stress, cancellous bone stress, cement stress, contact pressure between implants and micromotion between implants. However, bodyweight affects the cortical and cancellous bone stress more (stiff rise of the bone stresses; nonlinear relation) as compared to other output parameters (mostly linear relation). Comparing cortical and cancellous bone stress, the stress versus bodyweight curve is much stiffer (stiff rise in the curve) for cortical bone than cancellous bone and that even further increases as bone quality decreases. Especially considering obesity cases or obese people (very high bodyweight), the performance of the cemented acetabular component is poor. Graphical [ ]Keywords
This publication has 53 references indexed in Scilit:
- Experimental Validation of Finite Element Models of Intact and Implanted Composite Hemipelvises Using Digital Image CorrelationJournal of Biomechanical Engineering, 2012
- The effect of cup inclination and wear on the contact mechanics and cement fixation for ultra high molecular weight polyethylene total hip replacementsMedical Engineering & Physics, 2011
- In Vitro Fatigue Failure of Cemented Acetabular Replacements: A Hip Simulator StudyJournal of Biomechanical Engineering, 2008
- Obesity is a Major Risk Factor for Prosthetic Infection after Primary Hip ArthroplastyClinical Orthopaedics and Related Research, 2008
- Determination of interfacial fracture toughness of bone–cement interface using sandwich Brazilian disksEngineering Fracture Mechanics, 2007
- Fatigue failure in the cement mantle of a simplified acetabular replacement modelInternational Journal of Fatigue, 2007
- Trabecular bone modulus–density relationships depend on anatomic siteJournal of Biomechanics, 2003
- The influence of contact stress on the wear of UHMWPE for total replacement hip prosthesesWear, 1995
- Mechanical and textural properties of pelvic trabecular boneJournal of Biomechanics, 1993
- Stress distributions in the acetabular region—II. Effects of cement thickness and metal backing of the total hip acetabular componentJournal of Biomechanics, 1982