Abstract
Passive magnetic localization is a feasible method to perform a small-scale localization for robotics and biomedical devices, for instance, wireless capsule endoscopy. Suffering from the magnetic noises such as geomagnetic noise and magnetic disturbance in the environment, the localization accuracy of the passive magnetic localization is drastically influenced without a careful system design. In this paper, the signal-noise-ratio (SNR) of each sensor measurement is defined and applied in the design of both the localization algorithm and the measurement system. Simulations are conducted to verify the relationship between the measurement SNR and the accuracy of the distance estimation. Besides, strategies on selection of the target magnet, effective SNR threshold of the localization algorithm, and the optimal sensors distribution are proposed based on the SNR evaluation results. Extended experimental results show that by setting the SNR threshold at 6 dB and adopting the optimal sensors distribution, the root-mean-square localization error is improved from 0.0529 mm to 0.0364 mm and the orientation error is improved from 17.3° to 13.7° compared to the uniform sensors distribution. The proposed algorithm can be used for accurate localization and tracking for untethered devices or biomedical devices, for example, wireless capsule endoscopy.
Funding Information
  • Natural Science Foundation of Jiangsu Province, China (BK20191184)
  • Suzhou Science and Technology Council, China (SYG201839)

This publication has 20 references indexed in Scilit: