Abstract
The calibration of a tri-axial magnetometer (TAM) relies on a high-precision reference magnetic source that provides a known magnetic field, and its accuracy directly determines the final calibration precision. To address the challenge of calibrating TAMs across widely varying measurement ranges found in diverse applications, this paper proposes a coil design method featuring an adjustable coil constant (ACC). By integrating the target field method (TFM) and the non-dominated sorting genetic algorithm II with adaptive penalties (AP-NSGA-II), the coil structure is optimized to generate controllable magnetic fields spanning multiple orders of magnitude. This design enables the generation of a highly uniform magnetic field over a large central volume without relying on current sources with multiple output ranges. Experimental results demonstrate that the maximum magnetic field non-uniformity is 0.433% across different ACC settings, aligning well with design expectations. Furthermore, by combining the proposed system with the differential evolution (DE) algorithm, TAM calibration was performed at magnetic field intensities of 30,000 nT and 100 nT, achieving post-calibration residuals of 2.32 nT and 1.97 nT, respectively. Experiments verify that the coils designed with this method can provide high-precision, uniform reference magnetic fields for calibrating different types of TAMs across wide measurement ranges.
Funding Information
  • National Natural Science Foundation of China (52525501)
  • Quantum Science and Technology-National Science and Technology Major Project (2021ZD0300501)
  • National Natural Science Foundation of China (2025M773624)

This publication has 26 references indexed in Scilit: