Design of Uniform-Field Coils With Adjustable Coil Constant for Calibration of Triaxial Magnetometers Across Different Magnetic Field Ranges
- 16 April 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 75 (00189456) , 1-11
Abstract
The calibration of a tri-axial magnetometer (TAM) relies on a high-precision reference magnetic source that provides a known magnetic field, and its accuracy directly determines the final calibration precision. To address the challenge of calibrating TAMs across widely varying measurement ranges found in diverse applications, this paper proposes a coil design method featuring an adjustable coil constant (ACC). By integrating the target field method (TFM) and the non-dominated sorting genetic algorithm II with adaptive penalties (AP-NSGA-II), the coil structure is optimized to generate controllable magnetic fields spanning multiple orders of magnitude. This design enables the generation of a highly uniform magnetic field over a large central volume without relying on current sources with multiple output ranges. Experimental results demonstrate that the maximum magnetic field non-uniformity is 0.433% across different ACC settings, aligning well with design expectations. Furthermore, by combining the proposed system with the differential evolution (DE) algorithm, TAM calibration was performed at magnetic field intensities of 30,000 nT and 100 nT, achieving post-calibration residuals of 2.32 nT and 1.97 nT, respectively. Experiments verify that the coils designed with this method can provide high-precision, uniform reference magnetic fields for calibrating different types of TAMs across wide measurement ranges.Keywords
Funding Information
- National Natural Science Foundation of China (52525501)
- Quantum Science and Technology-National Science and Technology Major Project (2021ZD0300501)
- National Natural Science Foundation of China (2025M773624)
This publication has 26 references indexed in Scilit:
- Uniform bi-planar coil design considering coupling effect of finite thickness for SERF magnetometer array calibration inside magnetically shielded roomMeasurement, 2025
- Bandwidth compensation for ultra-high-sensitivity SERF magnetometers in magnetocardiac sensingMeasurement, 2024
- In-flight vector magnetometer calibration for FY-3E satelliteScience China Technological Sciences, 2023
- New Limit on the Permanent Electric Dipole Moment of Using Comagnetometry and SQUID DetectionPhysical Review Letters, 2019
- Three-Dimensional Magnetometer Calibration With Small Space Coverage for PedestriansIEEE Sensors Journal, 2014
- Magnetic anomaly map of the world: merging satellite, airborne, marine and ground-based magnetic data setsEarth and Planetary Science Letters, 2007
- Vectorial calibration of 3-D magnetic field sensor arrays2002
- TWOSTEP: A Fast Robust Algorithm for Attitude-Independent Magnetometer-Bias DeterminationThe Journal of the Astronautical Sciences, 2002
- A fast and elitist multiobjective genetic algorithm: NSGA-IIIEEE Transactions on Evolutionary Computation, 2002
- A target field approach to optimal coil designJournal of Physics D: Applied Physics, 1986