Optimized MLAA for quantitative non-TOF PET/MR of the brain
Open Access
- 2 December 2016
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 61 (24) , 8854-8874
Abstract
For quantitative tracer distribution in positron emission tomography, attenuation correction is essential. In a hybrid PET/CT system the CT images serve as a basis for generation of the attenuation map, but in PET/MR, the MR images do not have a similarly simple relationship with the attenuation map. Hence attenuation correction in PET/MR systems is more challenging. Typically either of two MR sequences are used: the Dixon or the ultra-short time echo (UTE) techniques. However these sequences have some well-known limitations. In this study, a reconstruction technique based on a modified and optimized non-TOF MLAA is proposed for PET/MR brain imaging. The idea is to tune the parameters of the MLTR applying some information from an attenuation image computed from the UTE sequences and a T1w MR image. In this MLTR algorithm, an parameter is introduced and optimized in order to drive the algorithm to a final attenuation map most consistent with the emission data. Because the non-TOF MLAA is used, a technique to reduce the cross-talk effect is proposed. In this study, the proposed algorithm is compared to the common reconstruction methods such as OSEM using a CT attenuation map, considered as the reference, and OSEM using the Dixon and UTE attenuation maps. To show the robustness and the reproducibility of the proposed algorithm, a set of 204 [18F]FDG patients, 35 [11C]PiB patients and 1 [18F]FET patient are used. The results show that by choosing an optimized value of in MLTR, the proposed algorithm improves the results compared to the standard MR-based attenuation correction methods (i.e. OSEM using the Dixon or the UTE attenuation maps), and the cross-talk and the scale problem are limited.Keywords
This publication has 38 references indexed in Scilit:
- Effect of scatter correction when comparing attenuation maps: Application to brain PET/MR2014
- Scattered PET data for attenuation‐map reconstruction in PET/MRIMedical Physics, 2014
- Attenuation Correction Synthesis for Hybrid PET-MR Scanners: Application to Brain StudiesIEEE Transactions on Medical Imaging, 2014
- Transmission-less attenuation correction in time-of-flight PET: analysis of a discrete iterative algorithmPhysics in Medicine & Biology, 2014
- A comparison of CT- and MR-based attenuation correction in neurological PETEuropean Journal of Nuclear Medicine and Molecular Imaging, 2014
- Combined PET/MR imaging in neurology: MR-based attenuation correction implies a strong spatial bias when ignoring boneNeuroImage, 2014
- Time-of-flight PET data determine the attenuation sinogram up to a constantPhysics in Medicine & Biology, 2012
- Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR ScannerJournal of Nuclear Medicine, 2011
- Attenuation correction in PET using consistency conditions and a three-dimensional templateIEEE Transactions on Nuclear Science, 2001
- A New Approach to the Emission Computerized Tomography Problem: Simultaneous Calculation of Attenuation and Activity CoefficientsIEEE Transactions on Nuclear Science, 1979