Clinical impact of using the deterministic patient dose calculation algorithm Acuros XB for lung stereotactic body radiation therapy
Authors:
- Zoann NugentZNZoann Nugent
- Ravinder ClaytonRCRavinder Clayton
- Peter DunscombePDPeter Dunscombe
- Harold LauHLHarold Lau
- Rao KhanRKRao Khan
- 19 August 2013
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Oncologica
- Vol. 53 (3) , 324-329
Abstract
Purpose. To evaluate the clinical impact of using the deterministic dose calculation algorithm, Acuros XB, for early stage lung cancer patients undergoing stereotactic body radiotherapy (SBRT). Material and methods. Seventy-seven stage I non-small cell lung cancer patients who underwent lung SBRT from 2008 to 2012 at our center were included in this study. All treatment plans originally calculated by the anisotropic analytical algorithm (AAA) were recalculated using the AAA and Acuros XB algorithms with identical monitor units and beam arrangements. The dose, dose distribution, conformality number (CN) and heterogeneity index (HI) of the target were determined for each plan. A paired matched t-test was used to evaluate the difference between the mean dose, the dose distribution, and the CN and HI for the target. The importance of tumor (volume, location), patient (pulmonary functional, body mass index) and treatment (number of SBRT beams) on the dose distributions obtained from the two algorithms was statistically determined using linear regression analyses. Results. The mean target dose was same for both algorithms. Compared to AAA, a small and significant difference in dose distribution in the target was found for the Acuros XB algorithm, resulting in lower conformity (−2.1%, p < 0.0001) and higher heterogeneity (p < 0.0001) of dose. Single logistic regression identified pulmonary function, number of beams and target location as being correlated with the difference of CN between the two calculations. Multivariate analysis indicated that the patient's pulmonary function (p = 0.0296) was the only predictor for the difference in conformality between the two dose calculation algorithms. Conclusions. In lung SBRT, the patient's pulmonary function is responsible for the difference in target dose distribution between the Acuros XB and AAA algorithms. The Acuros XB algorithm could be used to advantage in patients with compromised pulmonary function based on its accurate modeling of lung tissue in comparison to AAA.Keywords
This publication has 23 references indexed in Scilit:
- The influence of target and patient characteristics on the volume obtained from cone beam CT in lung stereotactic body radiation therapyRadiotherapy and Oncology, 2013
- Dosimetric evaluation of Acuros XB dose calculation algorithm with measurements in predicting doses beyond different air gap thickness for smaller and larger field sizesJournal of Medical Physics, 2013
- Critical Appraisal of Acuros XB and Anisotropic Analytic Algorithm Dose Calculation in Advanced Non-Small-Cell Lung Cancer TreatmentsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Rola tomografii komputerowej w ilościowej ocenie rozedmy płucPolish Journal of Radiology, 2012
- Clinical validation of the Acuros XB photon dose calculation algorithm, a grid-based Boltzmann equation solverActa Oncologica, 2011
- Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous mediaRadiation Oncology, 2011
- Validation of a new grid-based Boltzmann equation solver for dose calculation in radiotherapy with photon beamsPhysics in Medicine & Biology, 2010
- A 3D pencil-beam-based superposition algorithm for photon dose calculation in heterogeneous mediaPhysics in Medicine & Biology, 2008
- Evaluation of the analytical anisotropic algorithm in an extreme water–lung interface phantom using Monte Carlo dose calculationsJournal of Applied Clinical Medical Physics, 2007
- Comparison of a finite-element multigroup discrete-ordinates code with Monte Carlo for radiotherapy calculationsPhysics in Medicine & Biology, 2006