Accuracy of an Automatic Patient-Positioning System Based on the Correlation of Two Edge Images in Radiotherapy
Authors:
- 2 February 2010
- journal article
- research article
- Published by Springer Nature in Journal of Digital Imaging
- Vol. 24 (2) , 322-330
Abstract
We have clinically evaluated the accuracy of an automatic patient-positioning system based on the image correlation of two edge images in radiotherapy. Ninety-six head & neck images from eight patients undergoing proton therapy were compared with a digitally reconstructed radiograph (DRR) of planning CT. Two edge images, a reference image and a test image, were extracted by applying a Canny edge detector algorithm to a DRR and a 2D X-ray image, respectively, of each patient before positioning. In a simulation using a humanoid phantom, performed to verify the effectiveness of the proposed method, no registration errors were observed for given ranges of rotation, pitch, and translation in the x, y, and z directions. For real patients, however, there were discrepancies between the automatic positioning method and manual positioning by physicians or technicians. Using edged head coronal- and sagittal-view images, the average differences in registration between these two methods for the x, y, and z directions were 0.11 cm, 0.09 cm and 0.11 cm, respectively, whereas the maximum discrepancies were 0.34 cm, 0.38 cm, and 0.50 cm, respectively. For rotation and pitch, the average registration errors were 0.95° and 1.00°, respectively, and the maximum errors were 3.6° and 2.3°, respectively. The proposed automatic patient-positioning system based on edge image comparison was relatively accurate for head and neck patients. However, image deformation during treatment may render the automatic method less accurate, since the test image many differ significantly from the reference image.Keywords
This publication has 26 references indexed in Scilit:
- A dose‐volume‐based tool for evaluating and ranking IMRT treatment plansJournal of Applied Clinical Medical Physics, 2004
- Patient Set-Up Using Portal Images: 2D/2D Image Registration Using Mutual InformationComputer Aided Surgery, 2000
- Patient set-up using portal images: 2D/2D image registration using mutual informationComputer Aided Surgery, 2000
- Methods for fully automated verification of patient set-up in external beam radiotherapy with polygon shaped fieldsPhysics in Medicine & Biology, 1994
- Line Detection in Noisy and Structured Backgrounds Using Græco-Latin SquaresCVGIP: Graphical Models and Image Processing, 1993
- Edge detection in correlated noise using latin square masksPattern Recognition, 1988
- Compensating for heterogeneities in proton radiation therapyPhysics in Medicine & Biology, 1984
- Theory of edge detectionProceedings of the Royal Society of London. B. Biological Sciences, 1980
- Computer determination of the constituent structure of biological imagesComputers and Biomedical Research, 1971
- An Operator Which Locates Edges in Digitized PicturesJournal of the ACM, 1971