A miniature and mobile intermittent pneumatic compression device for the prevention of deep-vein thrombosis after joint replacement
Authors:
Open Access
- 1 January 2004
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 75 (5) , 584-587
Abstract
The WizAir-DVT is a miniature, lightweight (690 g), battery-operated and mobile intermittent pneumatic compression device (ICD), which enables continuous intraoperative use and immediate patient mobilization postoperatively. We compared its efficacy with a commonly used ICD, the Kendall SCD. Peak femoral vein flow velocity was measured in 20 apparently healthy volunteers at rest and with each device: we found no significant differences between them. A second prospective, randomized, clinical trial was used to compare the efficiency of the device in preventing deep venous thrombosis (DVT) after joint replacement in 50 patients (n=25/group). None developed DVT. Doppler ultrasonography revealed no significant differences. The WizAir-DVT antithrombotic compression device is as safe and effective as the Kendall SCD.Keywords
This publication has 5 references indexed in Scilit:
- Randomized clinical trial of intermittent pneumatic compression and low molecular weight heparin in traumaBritish Journal of Surgery, 2003
- Optimum Intermittent Pneumatic Compression Stimulus for Lower-limb Venous EmptyingEuropean Journal of Vascular and Endovascular Surgery, 2000
- The Effects of External Compression on Venous Blood Flow and Tissue Deformation in the Lower LegJournal of Biomechanical Engineering, 1999
- The Fibrinolytic Effects of Intermittent Pneumatic CompressionAnnals of Surgery, 1997
- Surgical Strategy for Management of Deep Venous Thrombosis of the Lower ExtremitiesWorld Journal of Surgery, 1996