A self-developed and constructed robot for minimally invasive cochlear implantation
- 4 March 2012
- journal article
- Published by Taylor & Francis in Acta Oto-Laryngologica
- Vol. 132 (4) , 355-360
Abstract
Conclusion: A robot built specifically for stereotactic cochlear implantation provides equal or better accuracy levels together with a better integration into a clinical environment, when compared with existing approaches based on industrial robots. Objectives: To evaluate the technical accuracy of a robotic system developed specifically for lateral skull base surgery in an experimental set-up reflecting the intended clinical application. The invasiveness of cochlear electrode implantation procedures may be reduced by replacing the traditional mastoidectomy with a small tunnel slightly larger in diameter than the electrode itself. Methods: The end-to-end accuracy of the robot system and associated image-guided procedure was evaluated on 15 temporal bones of whole head cadaver specimens. The main components of the procedure were as follows: reference screw placement, cone beam CT scan, computer-aided planning, pair-point matching of the surgical plan, robotic drilling of the direct access tunnel, and postoperative cone beam CT scan for accuracy assessment. Results: The mean accuracy at the target point (round window) was 0.56 ± 0.41 mm with an angular misalignment of 0.88 ± 0.40°. The procedural time for the registration process through the completion of the drilling procedure was 25 ± 11 min. The robot was fully operational in a clinical environment.Keywords
This publication has 15 references indexed in Scilit:
- Computerassistierte Präzisionschirurgie am Ohrat - Automatisierungstechnik, 2011
- Percutaneous cochlear implant drilling via customized frames: An in vitro studyOtolaryngology–Head and Neck Surgery, 2010
- A robot-guided minimally invasive approach for cochlear implant surgery: preliminary results of a temporal bone studyInternational Journal of Computer Assisted Radiology and Surgery, 2009
- Customized, rapid-production microstereotactic table for surgical targeting: description of concept and in vitro validationInternational Journal of Computer Assisted Radiology and Surgery, 2009
- A True Minimally Invasive Approach for Cochlear ImplantationOtology & Neurotology, 2008
- Percutaneous Cochlear Access Using Bone-Mounted, Customized Drill GuidesOtology & Neurotology, 2007
- Increase of Accuracy in Intraoperative Navigation Through High-Resolution Flat-Panel Volume Computed TomographyOtology & Neurotology, 2007
- Navigation as a quality management tool in cochlear implant surgeryThe Journal of Laryngology & Otology, 2004
- Veria Operation UpdatedORL, 2002
- Registration of head volume images using implantable fiducial markersIEEE Transactions on Medical Imaging, 1997