Increase of Accuracy in Intraoperative Navigation Through High-Resolution Flat-Panel Volume Computed Tomography
Authors:
- Soenke H. BartlingSBSoenke H. Bartling
- Martin LeinungMLMartin Leinung
- Johannes GrauteJGJohannes Graute
- Thomas RodtTRThomas Rodt
- Christian DullinCDChristian Dullin
- Hartmut BeckerHBHartmut Becker
- Thomas LenarzTLThomas Lenarz
- Timo StoverTSTimo Stover
- Omid MajdaniOMOmid Majdani
- 1 January 2007
- journal article
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 28 (1) , 129-134
Abstract
High-resolution imaging, as provided by flat-panel-based volume computed tomography (fpVCT), could increase navigation accuracy and could therefore improve image-guided procedures or make novel navigated surgery concepts possible. Intraoperative navigation is an accepted tool in head and neck surgery. However, its use is limited in the lateral cranial base because of its low surgical accuracy. Surgical accuracy is substantially influenced by the resolution of the underlying data set. The fpVCT offers a resolution of nearly two times higher than multislice computed tomography (MSCT). Target registration error (TRE), as a measurement for surgical navigation accuracy, should decrease when navigation is based on fpVCT data sets. An acrylic glass phantom with 37 fiducial points was scanned in a current MSCT and in an experimental fpVCT. Both data sets were imported in an optical navigation system. Five fiducial points were used for registration, and seven points were used for measuring TRE. The distance between the indicated pointer tip and the corresponding fiducial point in data set was measured as TRE. Registration and TRE measurement were repeated five times for each computed tomographic data set. Average TREs were calculated, and results were compared using t-test. The average TRE using MSCT (0.82 mm [standard deviation, 0.35 mm]) was significantly higher than that using fpVCT (0.46 mm [standard deviation, 0.22 mm]) (p < 0.01). Submillimeter surgical navigation accuracy is possible using high-resolution fpVCT. This could be highly beneficial in cranial base surgery navigation.Keywords
This publication has 10 references indexed in Scilit:
- Flat-Panel Volume Computed Tomography for Cochlear Implant Electrode Array Examination in Isolated Temporal Bone SpecimensOtology & Neurotology, 2006
- Image-guided surgery: what is the accuracy?Current Opinion in Otolaryngology & Head and Neck Surgery, 2005
- Sources of Registration Error with Image Guidance Systems During Endoscopic Anterior Cranial Base SurgeryOtolaryngology–Head and Neck Surgery, 2004
- Navigiert-kontrollierte KochleostomieHNO, 2004
- Der Einsatz von Flachbilddetektoren f r die CT-BildgebungDer Radiologe, 2003
- Image‐Guided Sinus Surgery: Lessons Learned from the First 1000 CasesOtolaryngology–Head and Neck Surgery, 2003
- Registration Accuracy and Practicability of Laser-Directed Surface MatchingComputer Aided Surgery, 2002
- Image-guided endoscopic ENT surgeryEuropean Archives of Oto-Rhino-Laryngology, 1997
- Computer‐Assisted Surgery in the Frontal and Maxillary SinusThe Laryngoscope, 1997
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984