A back-projection algorithm in the presence of an extra attenuating medium: towards EPID dosimetry for the MR-Linac
- 17 July 2017
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (15) , 6322-6340
Abstract
In external beam radiotherapy, electronic portal imaging devices (EPIDs) are frequently used for pre-treatment and for in vivo dose verification. Currently, various MR-guided radiotherapy systems are being developed and clinically implemented. Independent dosimetric verification is highly desirable. For this purpose we adapted our EPID-based dose verification system for use with the MR-Linac combination developed by Elekta in cooperation with UMC Utrecht and Philips. In this study we extended our back-projection method to cope with the presence of an extra attenuating medium between the patient and the EPID. Experiments were performed at a conventional linac, using an aluminum mock-up of the MRI scanner housing between the phantom and the EPID. For a 10 cm square field, the attenuation by the mock-up was 72%, while 16% of the remaining EPID signal resulted from scattered radiation. 58 IMRT fields were delivered to a 20 cm slab phantom with and without the mock-up. EPID reconstructed dose distributions were compared to planned dose distributions using the γ -evaluation method (global, 3%, 3 mm). In our adapted back-projection algorithm the averaged γ m e a n was 0.27 ± 0.06 , while in the conventional it was 0.28 ± 0.06 . Dose profiles of several square fields reconstructed with our adapted algorithm showed excellent agreement when compared to TPS.Keywords
Funding Information
- Elekta
This publication has 25 references indexed in Scilit:
- Feasibility of portal dosimetry for flattening filter‐free radiotherapyJournal of Applied Clinical Medical Physics, 2016
- The Australian Magnetic Resonance Imaging–Linac ProgramSeminars in Radiation Oncology, 2014
- Magnetic Resonance–Guided Adaptive Radiotherapy: A Solution to the FutureSeminars in Radiation Oncology, 2014
- The Rotating Biplanar Linac–Magnetic Resonance Imaging SystemSeminars in Radiation Oncology, 2014
- The Magnetic Resonance Imaging–Linac SystemSeminars in Radiation Oncology, 2014
- First MR images obtained during megavoltage photon irradiation from a prototype integrated linac‐MR systemMedical Physics, 2009
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetryMedical Physics, 2002
- Extraction of primary signal from EPIDs using only forward convolutionMedical Physics, 1997
- Dosimetric characteristics of a liquid-filled electronic portal imaging deviceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995