Pacing parameters and success rates of permanent His-bundle pacing in patients with narrow QRS: a single-centre experience
Authors:
- 15 December 2018
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 21 (5) , 763-770
Abstract
High and unstable capture thresholds affect the success rate of permanent His-bundle pacing (HBP). We aimed to introduce the modified techniques during different periods and the corresponding success rate of HBP implantation. Patients from a single centre who had intrinsic QRS < 120 ms and HBP attempts were included in the study. The success rate and pacing parameters were described for three periods based on procedural modifications, i.e. Stage 1 using the conventional HBP procedure (August 2012 to May 2013), Stage 2 with addition of the dual-lead method (June 2013 to October 2014), and Stage 3 with the further addition of stability assessment during fixation (November 2014 to October 2016). The patients with successful permanent HBP were followed. A total of 310 patients were included with the average age of 70.3 ± 10.7 years. The success rate of acute HBP was 84.85%, 98.3%, and 99.20% during Stages 1-3, respectively (P < 0.001). The permanent HBP implantation rates increased from 77.3% during Stage 1 to 85.7% during Stage 2 and 89.6% during Stage 3 (P = 0.07). The acute His-bundle capture threshold reduced from 1.30 ± 0.7 V/0.5 ms during Stage 1 to 1.11 ± 0.6 V/0.5 ms during Stage 2 and further to 0.85 ± 0.51 V/0.5 ms during Stage 3 (P < 0.001). At the 12-month follow-up, the mean change in the HBP threshold decreased from 0.60 ± 0.59 V/0.5 ms during Stage 1 to 0.33 ± 0.39 V/0.5 ms during Stage 3 (P = 0.002). The HBP implantation success rate, pacing threshold, and its stability during follow-up were improved by using the dual-lead method and stability assessment techniques.Keywords
Funding Information
- Science and Technology of Wenzhou (ZS2017010)
This publication has 14 references indexed in Scilit:
- A Novel Pacing Strategy With Low and Stable Output: Pacing the Left Bundle Branch Immediately Beyond the Conduction BlockCanadian Journal of Cardiology, 2017
- Benefits of Permanent His Bundle Pacing Combined With Atrioventricular Node Ablation in Atrial Fibrillation Patients With Heart Failure With Both Preserved and Reduced Left Ventricular Ejection FractionJournal of the American Heart Association, 2017
- Step-wise Approach to Permanent His Bundle PacingJournal of Innovations in Cardiac Rhythm Management, 2016
- Pacing and sensing optimization of permanent His-bundle pacing in cardiac resynchronization therapy/implantable cardioverter defibrillators patients: value of integrated bipolar configurationEP Europace, 2015
- Permanent His-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practiceHeart Rhythm, 2015
- Acute His‐Bundle Injury Current during Permanent His‐Bundle Pacing Predicts Excellent Pacing OutcomesPacing and Clinical Electrophysiology, 2015
- Hisian pacing restores cardiac functionJournal of Electrocardiology, 2013
- Safety and Performance of a System Specifically Designed for Selective Site PacingPacing and Clinical Electrophysiology, 2010
- A Feasible Approach for Direct His‐Bundle Pacing Using a New Steerable Catheter to Facilitate Precise Lead PlacementJournal of Cardiovascular Electrophysiology, 2006