On the use of a single‐fiber multipoint plastic scintillation detector for 192Ir high‐dose‐rate brachytherapy
- 9 May 2013
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 40 (6Part1) , 062101
Abstract
Purpose: The goal of this study was to prove the feasibility of using a single‐fiber multipoint plastic scintillation detector (mPSD) as anin vivo verification tool during 192Ir high‐dose‐rate brachytherapy treatments.Methods: A three‐point detector was built and inserted inside a catheter‐positioning template placed in a water phantom. A hyperspectral approach was implemented to discriminate the different optical signals composing the light output at the exit of the single collection optical fiber. The mPSD was tested with different source‐to‐detector positions, ranging from 1 to 5 cm radially and over 10.5 cm along the longitudinal axis of the detector, and with various integration times. Several strategies for improving the accuracy of the detector were investigated. The device's accuracy in detecting source position was also tested.Results: Good agreement with the expected doses was obtained for all of the scintillating elements, with average relative differences from the expected values of 3.4 ± 2.1%, 3.0 ± 0.7%, and 4.5 ± 1.0% for scintillating elements from the distal to the proximal. A dose threshold of 3 cGy improved the general accuracy of the detector. An integration time of 3 s offered a good trade‐off between precision and temporal resolution. Finally, the mPSD measured the radioactive source positioning uncertainty to be no more than 0.32 ± 0.06 mm. The accuracy and precision of the detector were improved by a dose‐weighted function combining the three measurement points and known details about the geometry of the detector construction.Conclusions: The use of a mPSD for high‐dose‐rate brachytherapy dosimetry is feasible. This detector shows great promise for development ofin vivo applications for real‐time verification of treatment delivery.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (262105)
- National Cancer Institute (1R01CA120198-01A2)
- Natural Sciences and Engineering Research Council of Canada (262105)
- National Cancer Institute (1R01CA120198-01A2)
This publication has 34 references indexed in Scilit:
- Solid state TL detectors for in vivo dosimetry in brachytherapyApplied Radiation and Isotopes, 2012
- In vivodosimetry with optically stimulated luminescent dosimeters, OSLDs, compared to diodes; the effects of buildup cap thickness and fabrication materialMedical Physics, 2011
- In vivo assessment of catheter positioning accuracy and prolonged irradiation time on liver tolerance dose after single-fraction 192Ir high-dose-rate brachytherapyRadiation Oncology, 2011
- In vivo real‐time dosimetric verification in high dose rate prostate brachytherapyMedical Physics, 2011
- In Vivo Dosimetry With a Linear MOSFET Array to Evaluate the Urethra Dose During Permanent Implant Brachytherapy Using Iodine-125International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Time‐resolved in vivo luminescence dosimetry for online error detection in pulsed dose‐rate brachytherapyMedical Physics, 2009
- Evaluation of a novel 4D in vivo dosimetry systemMedical Physics, 2009
- Characterization of a fiber‐coupled luminescence dosimetry system for online in vivo dose verification during brachytherapyMedical Physics, 2009
- In vivo thermoluminescence dosimetry dose verification of transperineal 192Ir high-dose-rate brachytherapy using CT-based planning for the treatment of prostate cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2003
- In vivourethral dose measurements: A method to verify high dose rate prostate treatmentsMedical Physics, 2000