In vivodosimetry with optically stimulated luminescent dosimeters, OSLDs, compared to diodes; the effects of buildup cap thickness and fabrication material
Authors:
- Paul A. JursinicPJPaul A. Jursinic
- Clifford J. YahnkeCYClifford J. Yahnke
- 16 September 2011
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 38 (10) , 5432-5440
Abstract
For external beam in vivo measurements, the dosimeter is normally placed on the patient's skin, and the dose to a point of interest inside the patient is derived from surface measurements. In order to obtain accurate and reliable measurements, which correlate with the dose values predicted by a treatment planning system, a dosimeter needs to be at a point of electronic equilibrium. This equilibrium is accomplished by adding material (buildup) above the detector. This paper examines the use of buildup caps in a clinical setting for two common detector types: OSLDs and diodes. Clinically built buildup-caps and commercially available hemispherical caps are investigated. The effects of buildup cap thickness and fabrication material on field-size correction factors, C(FS), are reported, and differences between the effects of thickness and fabrication material are explained based on physical parameters. Measurements are made on solid water phantoms for 6 and 15 MV x-ray beams. Two types of dosimeters are used: OSLDs, InLight∕OSL Nanodot dosimeters (Landauer, Inc., Glenwood, IL) and a P-type surface diode (Standard Imaging, Madison, WI). Buildup caps for these detectors were fabricated out of M3, a water-equivalent material, and sheet-metal stock of Al, Cu, and Pb. Also, commercially available hemispherical buildup caps made of plastic water and brass (Landauer, Inc., Glenwood, IL) were used with Nanodots. OSLDs were read with an InLight microStar reader (Landauer, Inc., Glenwood, IL). Dose calculations were carried out with the XiO treatment planning system (CMS∕Elekta, Stockholm) with tissue heterogeneity corrections. For OSLDs and diodes, when measurements are made with no buildup cap a change in C(FS) of 200% occurs for a field-size change from 3 cm × 3 cm to 30 cm × 30 cm. The change in C(FS) is reduced to about 4% when a buildup cap with wall thickness equal to the depth of maximum dose is used. Buildup caps with larger wall thickness do not cause further reduction in C(FS). The buildup cap fabrication material has little or no effect on C(FS). The perturbation to the delivered dose caused by placing a detector with a buildup cap on the surface of a patient is measured to be 4%-7%. A comparison between calculated dose and dose measured with a Nanodot and a diode for 6 and 15 MV x-rays is made. When C(FS) factors are carefully determined and applied to measurements made on a phantom, the differences between measured and calculated doses were found to be between ±1.3%. OSLDs and diodes with appropriate buildup caps can be used to measure dose on the surface of a patient and predict the delivered dose to depth dmax in a range of ±1.3% for 100 cGy. The buildup cap: can be fabricated from any material examined in this work, is best with wall thickness dmax, and causes a perturbation to the delivered dose of 4%-7% when the wall thickness is dmax. OSLDs and diodes with buildup caps can both give accurate measurements of delivered dose.Keywords
This publication has 44 references indexed in Scilit:
- Optically stimulated luminescence (OSL) dosimetry in medicinePhysics in Medicine & Biology, 2008
- Evaluation of optically stimulated luminescence (OSL) dosimeters for passive dosimetry of high‐energy photon and electron beams in radiotherapyMedical Physics, 2007
- Technical performance of the Luxel Al2O3:C optically stimulated luminescence dosemeter element at radiation oncology and nuclear accident dose levelsRadiation Protection Dosimetry, 2006
- In vivo dosimetry during conformal radiotherapyRadiotherapy and Oncology, 1999
- Evaluation of a dual bias dual metal oxide‐silicon semiconductor field effect transistor detector as radiation dosimeterMedical Physics, 1994
- On the cause of the variation in tissue-maximum ratio values with source-to-detector distanceMedical Physics, 1993
- Comparison of entrance and exit dose measurements using ionization chambers and silicon diodesPhysics in Medicine & Biology, 1991
- Characteristics of silicon diodes as patient dosemeters in external radiation therapyRadiotherapy and Oncology, 1988
- Effects of radiation damage on p-type silicon detectorsPhysics in Medicine & Biology, 1983
- Electron contamination of a high-energy X-ray beamPhysics in Medicine & Biology, 1978