Optimizing the design of bipolar nerve cuff electrodes for improved recording of peripheral nerve activity
- 30 March 2017
- journal article
- research article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 14 (3) , 036015
Abstract
Objective. Differential measurement of efferent and afferent peripheral nerve activity offers a promising means of improving the clinical utility of implantable neuroprostheses. The tripolar nerve cuff electrode has historically served as the gold standard for achieving high signal-to-noise ratios (SNRs) of the recordings. However, the symmetrical geometry of this electrode array (i.e. electrically-shorted side contacts) precludes it from measuring electrical signals that can be used to obtain directional information. In this study, we investigated the feasibility of using a bipolar nerve cuff electrode to achieve high-SNR of peripheral nerve activity. Approach. A finite element model was implemented to investigate the effects of electrode design parameters—electrode length, electrode edge length (EEL), and a conductive shielding layer (CSL)—on simulated single fiber action potentials (SFAP) and also artifact noise signals (ANS). Main results. Our model revealed that the EEL was particularly effective in increasing the peak-to-peak amplitude of the SFAP (319%) and reducing the common mode ANS (67%) of the bipolar cuff electrode. By adding a CSL to the bipolar cuff electrode, the SNR was found to be 65.2% greater than that of a conventional tripolar cuff electrode. In vivo experiments in anesthetized rats confirmed that a bipolar cuff electrode can achieve a SNR that is 38% greater than that achieved by a conventional tripolar cuff electrode (p < 0.05). Significance. The current study showed that bipolar nerve cuff electrodes can be designed to achieve SNR levels that are comparable to that of tripolar configuration. Further work is needed to confirm that these bipolar design parameters can be used to record bi-directional neural activity in a physiological setting.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (06519)
- Canada Foundation for Innovation (31506)
This publication has 44 references indexed in Scilit:
- Effect on signal-to-noise ratio of splitting the continuous contacts of cuff electrodes into smaller recording areasJournal of NeuroEngineering and Rehabilitation, 2013
- Improvement of signal-to-interference ratio and signal-to-noise ratio in nerve cuff electrode systemsPhysiological Measurement, 2012
- Tripolar-cuff deviation from ideal model: Assessment by bioelectric field simulations and saline-bath experimentsMedical Engineering & Physics, 2007
- Selective Recording of the Canine Hypoglossal Nerve Using a Multicontact Flat Interface Nerve ElectrodeIEEE Transactions on Biomedical Engineering, 2005
- Skin contact forces extracted from human nerve signals-a possible feedback signal for FES-aided control of standingIEEE Transactions on Biomedical Engineering, 2003
- Artefact reduction with alternative cuff configurationsIEEE Transactions on Biomedical Engineering, 2003
- Modeling study of peripheral nerve recording selectivityIEEE Transactions on Rehabilitation Engineering, 2000
- Integrating Sensory Nerve Signals Into Neural Prosthesis DevicesNeuromodulation: Technology at the Neural Interface, 2000
- Conduction velocity changes along lumbar primary afferent fibers in catsExperimental Neurology, 1984
- Partition of Gross Peripheral Nerve Activity into Single Unit Responses by Correlation TechniquesScience, 1975