Late Loosening of Press-Fit Cementless Acetabular Components
Authors:
- W. A. JiranekWJW. A. Jiranek
- D. R. WhiddonDWD. R. Whiddon
- W. T. JohnstoneWJW. T. Johnstone
- 1 January 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 418 (418) , 172-178
Abstract
Between January 1987 and September 1990, 67 consecutive cementless total hip arthroplasties were implanted in 59 patients by one surgeon. The mean age of the patients at surgery was 57 years (range, 23–80 years). All acetabular components were plasma-sprayed titanium hemispheric cups with four peripheral rim fins, but additional screw fixation was not used. The mean followup was 10.4 years (range, 8.8–12.5 years). With revision as the endpoint, the failure rate of this acetabular component at a mean of 10.4 years was 28% (19/67). Of the 56 patients (56 hips) with radiographic followup, loosening of the acetabular shell occurred in 10 hips in 10 patients (18%). Seventy percent of these loosened cups failed by tilt which occurred in a rapid manner; all of the patients required revision surgery. We examined the manner of loosening of a press-fit acetabular component after early (5-year) results showed high hip scores and a low failure rate.Keywords
This publication has 28 references indexed in Scilit:
- Tapered Titanium Cementless Total Hip ReplacementsClinical Orthopaedics and Related Research, 2001
- Isolated revision acetabuloplasty using a porous-coated cementless acetabular component without removal of a well-fixed femoral component: A 3- to 9-year follow-up studyThe Journal of Arthroplasty, 1997
- Cementless hemispheric acetabular componentsThe Journal of Arthroplasty, 1996
- Failed innovations—no concern of ours?Acta Orthopaedica, 1996
- Comparison of the results of cemented, porous-ingrowth, and threaded acetabular cup fixationThe Journal of Arthroplasty, 1995
- The Harris-Galante porous acetabular component press-fit without screw fixationThe Journal of Arthroplasty, 1994
- Cementless hemispheric acetabular components: 2–4-year resultsThe Journal of Arthroplasty, 1993
- Structural and cellular assessment of bone quality of proximal femurBone, 1993
- Evaluation of cementless acetabular component migrationThe Journal of Arthroplasty, 1989
- Effects of prosthetic acetabular replacement on strains in the pelvisJournal of Orthopaedic Research, 1985