Repeatability of a Commercially Available Adaptive Optics Visual Simulator and Aberrometer in Normal and Keratoconic Eyes

Authors:
  • RSRohit Shetty
    Rohit Shetty
  • SKShruti Kochar
    Shruti Kochar
  • TGTushar Grover
    Tushar Grover
  • PKPooja Khamar
    Pooja Khamar
  • PKPallak Kusumgar
    Pallak Kusumgar
  • KSKanchan Sainani
    Kanchan Sainani

Abstract
Visual adaptive optics provided repeatable aberrometry data in both normal and keratoconic eyes. For most of the parameters, the repeatability in eyes with early keratoconus was somewhat better than that for normal eyes. The repeatability of the Zernike terms was acceptable for 3rd order (trefoil and coma) and spherical aberrations. Therefore, visual adaptive optics was a suitable tool to perform repeatable aberrometric measurements. [J Refract Surg. 2017;33(11):769-772.].