Repeatability of a Commercially Available Adaptive Optics Visual Simulator and Aberrometer in Normal and Keratoconic Eyes
Authors:
- Rohit ShettyRSRohit Shetty
- Shruti KocharSKShruti Kochar
- Tushar GroverTGTushar Grover
- Pooja KhamarPKPooja Khamar
- Pallak KusumgarPKPallak Kusumgar
- Kanchan SainaniKSKanchan Sainani
- Abhijit Sinha RoyARAbhijit Sinha Roy
- 1 November 2017
- journal article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 33 (11) , 769-772
Abstract
Visual adaptive optics provided repeatable aberrometry data in both normal and keratoconic eyes. For most of the parameters, the repeatability in eyes with early keratoconus was somewhat better than that for normal eyes. The repeatability of the Zernike terms was acceptable for 3rd order (trefoil and coma) and spherical aberrations. Therefore, visual adaptive optics was a suitable tool to perform repeatable aberrometric measurements. [J Refract Surg. 2017;33(11):769-772.].Keywords
This publication has 26 references indexed in Scilit:
- Adaptive optics for studying visual function: A comprehensive reviewJournal of Vision, 2011
- Evaluation of the Comparability and Repeatability of Four Wavefront AberrometersInvestigative Ophthalmology & Visual Science, 2011
- Wave aberration of human eyes and new descriptors of image optical quality and visual performanceJournal of Cataract & Refractive Surgery, 2010
- Measurement of corneal aberrations for customisation of intraocular lens asphericity: impact on quality of vision after micro-incision cataract surgeryBritish Journal of Ophthalmology, 2009
- Customized hydrogel contact lenses for keratoconus incorporating correction for vertical coma aberrationOphthalmic and Physiological Optics, 2009
- Performance of Wavefront‐Guided Soft Lenses in Three Keratoconus SubjectsOptometry and Vision Science, 2008
- Wavefront-guided customized corneal ablationCurrent Opinion in Ophthalmology, 2008
- Analysis of the performance of the Hartmann–Shack sensor in the human eyeJournal of the Optical Society of America A, 2000
- Statistics Notes: Measurement errorBMJ, 1996
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986