Measurement uncertainties of scanning microwave radiometers and their influence on temperature profiling
Preprint
- 19 June 2023
- preprint
- Published by Copernicus GmbH in EGUsphere
- Vol. 2023, 1-24
Abstract
In order to improve observations of the atmospheric boundary layer (ABL), Europe's network of meteorological and hydrological services EUMETNET as well as the European Research Infrastructure ACTRIS are currently working on building networks of microwave radiometers (MWRs). Elevation-scanning MWRs are well suited to obtain temperature profiles of the atmosphere, especially within the ABL. Understanding and assessing measurement uncertainties of scanning MWRs is therefore crucial for accurate temperature profiling. In this paper we discuss measurement uncertainties due to the instrument set-up and originating from external sources, namely (1) horizontal inhomogeneities of the atmosphere, (2) pointing errors or a tilt of the instrument, (3) physical obstacles in line of sight of the instrument, and (4) radio frequency interference (RFI). Horizontal inhomogeneities from observations at the Jülich Observatory of Cloud Evolution (JOYCE) are shown to have a small impact on retrieved temperature profiles (< |0.22 K| in the 25th/75th percentiles below 3000 m). Typical instrument tilts, that could be caused by uncertainties during the instrument set-up, also have a very small impact on temperature profiles and are smaller than 0.1 K below 3000 m for up to 1° of tilt. Physical obstacles at ambient temperatures and in line of sight and filling the complete beam of the MWR at the lowest elevation angle of 5.4° have to be at least 600 m away from the instrument in order to have an impact of less than 0.1 K on obtained temperature profiles. If the obstacle is 5 K warmer than its surroundings then the obstacle should be at least 2700 m away. Finally, we present an approach on how to detect RFI with an MWR with azimuth and elevation scanning capabilities. In this study we detect RFIs in a water vapor channel that does not influence the temperature retrieval, but would be relevant if the MWR were used to detect horizontal humidity inhomogeneities.Keywords
All Related Versions
- Published version: Atmospheric Measurement Techniques, 17 (1), 219.
This publication has 19 references indexed in Scilit:
- EUMETNET opens to microwave radiometers for operational thermodynamical profiling in Europe2021
- Statistical retrieval of thin liquid cloud microphysical properties using ground‐based infrared and microwave observationsJournal of Geophysical Research: Atmospheres, 2016
- Assimilation of humidity and temperature observations retrieved from ground‐based microwave radiometers into a convective‐scale NWP modelQuarterly Journal of the Royal Meteorological Society, 2016
- Calibrating ground‐based microwave radiometers: Uncertainty and driftsRadio Science, 2016
- Accuracy of Boundary Layer Temperature Profiles Retrieved With Multifrequency Multiangle Microwave RadiometryIEEE Transactions on Geoscience and Remote Sensing, 2007
- A network suitable microwave radiometer for operational monitoring of the cloudy atmosphereAtmospheric Research, 2005
- The effect of the half-width of the 22-GHz water vapor line on retrievals of temperature and water vapor profiles with a 12-channel microwave radiometerIEEE Transactions on Geoscience and Remote Sensing, 2005
- Accuracy of cloud liquid water path from ground‐based microwave radiometry 1. Dependency on cloud model statisticsRadio Science, 2003
- Accuracy of cloud liquid water path from ground‐based microwave radiometry 2. Sensor accuracy and synergyRadio Science, 2003
- Water vapor microwave continuum absorption: A comparison of measurements and modelsRadio Science, 1998