Development of a Redox-Polymer-Based Electrochemical Glucose Biosensor Suitable for Integration in Microfluidic 3D Cell Culture Systems
Authors:
- L. Navarro-NaterasLNL. Navarro-Nateras
- Jancarlo Diaz-GonzalezJDJancarlo Diaz-Gonzalez
- Diana Aguas-ChantesDADiana Aguas-Chantes
- Lucy L. Coria-OriundoLCLucy L. Coria-Oriundo
- Fernando BattagliniFBFernando Battaglini
- José Luis Ventura-GallegosJVJosé Luis Ventura-Gallegos
- Alejandro Zentella-DehesaAZAlejandro Zentella-Dehesa
- Goldie OzaGoldie Oza
- L. G. ArriagaLAL. G. Arriaga
- Jannu R. Casanova-MorenoJCJannu R. Casanova-Moreno
Open Access
- 27 May 2023
- journal article
- research article
- Published by MDPI AG in Biosensors
- Vol. 13 (6) , 582
Abstract
The inclusion of online, in situ biosensors in microfluidic cell cultures is important to monitor and characterize a physiologically mimicking environment. This work presents the performance of second-generation electrochemical enzymatic biosensors to detect glucose in cell culture media. Glutaraldehyde and ethylene glycol diglycidyl ether (EGDGE) were tested as cross-linkers to immobilize glucose oxidase and an osmium-modified redox polymer on the surface of carbon electrodes. Tests employing screen printed electrodes showed adequate performance in a Roswell Park Memorial Institute (RPMI-1640) media spiked with fetal bovine serum (FBS). Comparable first-generation sensors were shown to be heavily affected by complex biological media. This difference is explained in terms of the respective charge transfer mechanisms. Under the tested conditions, electron hopping between Os redox centers was less vulnerable than H2O2 diffusion to biofouling by the substances present in the cell culture matrix. By employing pencil leads as electrodes, the incorporation of these electrodes in a polydimethylsiloxane (PDMS) microfluidic channel was achieved simply and at a low cost. Under flow conditions, electrodes fabricated using EGDGE presented the best performance with a limit of detection of 0.5 mM, a linear range up to 10 mM, and a sensitivity of 4.69 μA mM−1 cm−2.Keywords
Funding Information
- CIDETEQ (PI.230033)
This publication has 67 references indexed in Scilit:
- A Critical Review of Glucose Biosensors Based on Carbon Nanomaterials: Carbon Nanotubes and GrapheneSensors, 2012
- Dual-electrode microfluidic cell for characterizing electrocatalystsLab on a Chip, 2012
- A Fluorescence Method for Measurement of Glucose Transport in Kidney CellsDiabetes Technology & Therapeutics, 2011
- Shiga toxin glycosphingolipid receptors in microvascular and macrovascular endothelial cells: differential association with membrane lipid raft microdomainsJournal of Lipid Research, 2011
- How to embed three-dimensional flexible electrodes in microfluidic devices for cell culture applicationsLab on a Chip, 2011
- Development of High-Throughput Quantitative Assays for Glucose Uptake in Cancer Cell LinesMolecular Imaging and Biology, 2010
- A Tale of Two Compartments: Interstitial Versus Blood Glucose MonitoringDiabetes Technology & Therapeutics, 2009
- Fabrication of carbon microelectrodes with a micromolding technique and their use in microchip-based flow analysesThe Analyst, 2004
- Production of Adrenomedullin in Macrophage Cell Line and Peritoneal MacrophageJournal of Biological Chemistry, 1998
- The Electrochemistry Group Medal Lecture. Electron self-exchange dynamics between redox sites in polymersFaraday Discussions of the Chemical Society, 1989