An automated system for the functional evaluation of oscillometric non-invasive blood pressure monitors
Authors:
- J. N. AmooreJAJ. N. Amoore
- D. H. T. ScottDSD. H. T. Scott
- 1 January 1995
- journal article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 19 (5) , 162-176
Abstract
Most of the automated non-invasive blood pressure (NIBP) monitors in clinical use rely on the oscillometric technique for measuring the blood pressures. The functional evaluation of these monitors has relied on comparison with intra-arterial blood pressure measurements or comparison with ‘standard’ non-invasive blood pressure meters such as the Hawksley random zero sphygmomanometer. The commercial availability of simulated arm test instruments provides the opportunity for developing an alternative evaluation method. We describe the control of an NIBP simulator by a personal computer which is also linked to the NIBP monitor under test. An evaluation protocol is described and the data analysed to present graphically and in tabular form an assessment of the NIBP monitor.Keywords
This publication has 12 references indexed in Scilit:
- Survey of Automated Noninvasive Blood Pressure MonitorsJournal of Clinical Engineering, 1994
- Standard algorithm of blood-pressure measurement by the oscillometric methodMedical & Biological Engineering & Computing, 1994
- Comparison of oscillometric and intra-arterial blood pressure and pulse measurementJournal of Medical Engineering & Technology, 1994
- Noninvasive blood pressure measurements with single and twin‐hose systems—do mixtures matter?Anaesthesia, 1993
- Assessment of oscillometric non-invasive blood pressure monitors using the Dynatech Nevada Cuff Link analyserJournal of Medical Engineering & Technology, 1993
- Review Of Methods & Simulators For Evaluation Of Noninvasive Blood Pressure MonitorsJournal of Clinical Engineering, 1992
- Cardiac cycle phase uncertainty: another source of error in indirect blood pressure measurementJournal of Medical Engineering & Technology, 1992
- Clinical evaluation of the Copal UA-251 and the Dinamap 1848 automatic blood–pressure monitorsJournal of Medical Engineering & Technology, 1988
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983
- Clinical evaluation of Dinamap 845 automated blood pressure recorder.Heart, 1980