Count-rate management in 131I SPECT/CT calibration
Open Access
- 26 January 2025
- journal article
- research article
- Published by Springer Nature in EJNMMI Physics
- Vol. 12 (1) , 1-19
Abstract
Background: System calibration is essential for accurate SPECT/CT dosimetry. However, count losses due to dead time and pulse pileup may cause calibration errors, in particular for 131I, where high count rates may be encountered. Calibration at low count rates should also be avoided to minimise detrimental effects from e.g. background counts and statistical fluctuations. This paper aims to present experimental data illustrating count-rate dependencies and to propose practical routines to mitigate errors in the 131I calibration procedure without needing advanced analysis tools. Results: The sensitivities of two General Electric (GE) Discovery 670 Pro systems were assessed using two Jaszczak phantom geometries. SPECT/CT data were collected over two months, starting with an initial 131I content of > 2 GBq, decaying to approximately 20 MBq. This allowed for a detailed analysis of count losses due to dead time and pulse pileup. From the sensitivity analysis, it was shown that robust calibration was obtained for 131I phantom activities ranging between 250 and 1500 MBq. Conclusions: The results show that adequate corrections for dead-time and pulse-pileup counting losses are essential for accurate calibration. It is argued that loss corrections should be based on total spectrum count rates in projections and not only on the 364.5 keV energy window data. The measurement campaigns presented in this paper, using basic tools and equipment, may serve as a model for establishing routines for count-loss corrections as well as for system calibration and regular control of system sensitivity. The data suggest that analysis of source and count concentration in a homogeneous Jaszczak phantom offers robust calibration, whereas analysis of source strength and counts in a delineated phantom insert offers a practical and robust method for regular quality control.Keywords
Funding Information
- Karolinska Institute
This publication has 13 references indexed in Scilit:
- Quantitative SPECT (QSPECT) at high count rates with contemporary SPECT/CT systemsEJNMMI Physics, 2021
- Gamma detector dead time correction using Lambert W functionEJNMMI Physics, 2020
- EANM Dosimetry Committee series on standard operational procedures for internal dosimetry for 131I mIBG treatment of neuroendocrine tumoursEJNMMI Physics, 2020
- Comprehensive SPECT/CT system characterization and calibration for 177Lu quantitative SPECT (QSPECT) with dead-time correctionEJNMMI Physics, 2020
- Standardised quantitative radioiodine SPECT/CT Imaging for multicentre dosimetry trials in molecular radiotherapyPhysics in Medicine & Biology, 2019
- Accuracy and precision assessment for activity quantification in individualized dosimetry of 177Lu-DOTATATE therapyEJNMMI Physics, 2017
- MIRD Pamphlet No. 26: Joint EANM/MIRD Guidelines for Quantitative 177Lu SPECT Applied for Dosimetry of Radiopharmaceutical TherapyJournal of Nuclear Medicine, 2015
- MIRD Pamphlet No. 24: Guidelines for Quantitative 131I SPECT in Dosimetry ApplicationsJournal of Nuclear Medicine, 2013
- MIRD Pamphlet No. 23: Quantitative SPECT for Patient-Specific 3-Dimensional Dosimetry in Internal Radionuclide TherapyJournal of Nuclear Medicine, 2012
- Physics in Nuclear MedicineMedical Physics, 2004