Confocal micro-x-ray fluorescence spectrometer for light element analysis
Open Access
- 1 August 2012
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 83 (8) , 083703
Abstract
An existing micro-x-ray fluorescence (micro-XRF) spectrometer designed for light element analysis (6 ≤ Z ≤ 14) has been extended to confocal geometry: a second polycapillary x-ray optic has been introduced in front of the energy dispersive x-ray detector. New piezo positioners for optimum alignment of both optics have been installed inside the vacuum chamber. The spectrometer offers now the possibility of true 3D elemental analysis in the micrometer regime. Depth resolution varies between 100 μm at 1 keV fluorescence energy (Na-Kα) and 30 μm for 17.5 keV (Mo). To further extend analytical capabilities a second x-ray tube with a Rh anode has been acquired to supplement to existing Mo anode tube. Lower limits of detection have been determined to be in the ppm region for confocal geometry. The spectrometer has been characterized and tested using different samples. Furthermore, results have been compared with SR micro-XRF to show the capabilities and limitations of this spectrometer.Keywords
This publication has 15 references indexed in Scilit:
- Measurements of the Binding of a Large Protein Using a Substrate Density-Controlled DNA ChipAnalytical Chemistry, 2011
- Development of a new confocal 3D-XRF instrument with an X-ray tubeJournal of Analytical Atomic Spectrometry, 2010
- Improved micro x-ray fluorescence spectrometer for light element analysisReview of Scientific Instruments, 2010
- PART II (Portable ART analyzer)—development of a XRF spectrometer adapted for the study of artworks in the Kunsthistorisches Museum, ViennaX-Ray Spectrometry, 2009
- Applications of a new portable (micro) XRF instrument having low‐Z elements determination capability in the field of works of artX-Ray Spectrometry, 2008
- A portable micro-X-ray fluorescence spectrometer with polycapillary optics and vacuum chamber for archaeometric and other applicationsSpectrochimica Acta Part B: Atomic Spectroscopy, 2007
- Determination of the elemental distribution in human joint bones by SR micro XRFX-Ray Spectrometry, 2007
- Three-dimensional micro-XRF investigations of paint layers with a tabletop setupSpectrochimica Acta Part B: Atomic Spectroscopy, 2005
- Three-Dimensional Trace Element Analysis by Confocal X-ray Microfluorescence ImagingAnalytical Chemistry, 2004
- A new 3D micro X-ray fluorescence analysis set-up – First archaeometric applicationsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2003