A national dosimetry audit of intraoperative radiotherapy
Authors:
- D J EatonDED J Eaton
- B EarnerBEB Earner
- P FaulknerPFP Faulkner
- N DancerNDN Dancer
- 1 December 2013
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 86 (1032) , 20130447
Abstract
National dosimetry audits are a fundamental part of quality assurance in radiotherapy, especially for new techniques. Intraoperative radiotherapy with a compact mobile kilovoltage X-ray source is a novel approach for the treatment of breast and other cancers. All seven current clinical sites in the UK were audited by a single visiting group and set of measurement equipment. Measurements of output, isotropy and depth doses were performed using an ion chamber in solid water, thermoluminescent dosemeters and radiochromic film, respectively. The mean difference between measured and planned dose across all centres was -3.2±2.7%. Measured isotropy was within ±3% around the lateral plane of the X-ray source and +11±4% in the forward direction compared with the lateral plane. Measured depth doses were agreed within 5±2% of manufacturer-provided calibration values or a mean gamma index of 97% at a tolerance of 7%/0.5 mm. Agreement within measurement uncertainties was found for all three parameters except forward anisotropy, which is unlikely to be clinically significant. Steep dose gradients increase the sensitivity to small variations in positioning, but these tests are practical for use in interdepartmental audits and local baseline comparison. The first UK interdepartmental audit of intraoperative radiotherapy builds confidence in the delivery of this treatment.Keywords
This publication has 23 references indexed in Scilit:
- In vivo dosimetry with optically stimulated dosimeters and RTQA2 radiochromic film for intraoperative radiotherapy of the breastMedical Physics, 2013
- A regional audit of kilovoltage X-rays — a single centre approachBritish Journal of Radiology, 2008
- Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al–4 mm Cu HVL)Physics in Medicine & Biology, 2005
- Targeted intra-operative radiotherapy (Targit): An innovative method of treatment for early breast cancerAnnals of Oncology, 2001
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- A dosimetric intercomparison of electron beams in UK radiotherapy centresPhysics in Medicine & Biology, 1997
- The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al - 4 mm Cu HVL; 10 - 300 kV generating potential)Physics in Medicine & Biology, 1996
- Radiation properties of a miniature X-ray device for radiosurgeryBritish Journal of Radiology, 1996
- The development of an interdepartmental audit as part of a physics quality assurance programme for external beam therapyBritish Journal of Radiology, 1994
- A dosimetric intercomparison of megavoltage photon beams in UK radiotherapy centres.1992