Pressure Estimator of Airway Narrowing in Total Liquid Ventilation: First Results in Guinea Pigs

Authors:

Abstract
Total Liquid Ventilation (TLV) consists of ventilating the lungs of mammals with a breathable liquid (PFOB). This technology is being considered for extremely premature baby (less than 1 kg). However, liquid expiration in the high flexible airways can trigger a sudden tracheal collapse, observed as a sudden pressure drop at the endotracheal tube connector. To our knowledge, there is no reported TLV experiments in mammals of very low weight, ranged from 507 to 552 g (n=4), nor has tracheal collapse been observed. The objective is to develop a real-time pressure estimator to monitor the airway opening as well as to control the expiratory flow through a 2 mm diameter endotracheal tube. The experiments were carried out on 4 guinea pigs with a tracheotomy using a liquid ventilator during 4 hours. After 1 hour of TLV, acceptable blood gases were obtained proving that total liquid ventilation was tolerated. The experimental flow and pressure measurements from the first two experiments were used to learn the non-linear dynamics and the model parameters. The model was implemented in real-time on the microcontroller and used to test two control strategies of the expiratory flow. The last two guinea pigs were ventilated with the implemented estimator to detect sudden airway narrowing and to automatically control the expiratory flow rate. The results confirmed the effectiveness of the method in reversing the sudden pressure drop. In conclusion, the pressure estimator can be helpful in controlling liquid ventilation in mammals of very low weight, comparable to extreme premature babies.