Torsional fatigue and endurance limit of a size 30 .06 ProFile rotary instrument
Authors:
- 25 May 2004
- journal article
- Published by Wiley in International Endodontic Journal
- Vol. 37 (6) , 370-373
Abstract
Aim To evaluate the torsional cyclic fatigue characteristics and specifically the endurance limit (EL) of a nickel–titanium rotary instrument. Methodology Size 30 .06 taper ProFile instruments were evaluated. The equipment was assembled according to the ANSI/ADA Specification No. 28. The motor was programmed to repeatedly rotate to a selected deflection angle (DA) and then return to zero (cycle). Testing started at 200° and was continued at decreasing angles until 106 cycles were reached without instrument fracture. Ten instruments were tested at each DA. The mean log number of cycles to fracture and standard deviation were determined for each DA at which fracture occurred. The DA at which 106 cycles were reached without instrument fracture corresponded by definition to the EL. Analysis of variance and pairwise comparisons using Duncan's multiple range test were performed to detect significant differences among the mean log number of cycles of the different DA. Significance was determined at the 95% confidence level. Results Instruments cycled at larger DA consistently demonstrated fewer cycles to fracture than those cycled at smaller DA. The differences among the mean log number of cycles of the different DA were statistically significant (P < 0.001). Cycles of 106 were completed without instrument fracture at 2.5°. Conclusions A torsional fatigue profile was generated for a specific NiTi rotary instrument. The EL was 2.5°.Keywords
This publication has 11 references indexed in Scilit:
- ProTaper rotary root canal preparation: assessment of torque and force in relation to canal anatomyInternational Endodontic Journal, 2003
- Influence of Structure on Nickel-Titanium Endodontic Instruments FailureJournal of Endodontics, 2001
- Impact of Two Theoretical Cross-Sections on Torsional and Bending Stresses of Nickel-Titanium Root Canal Instrument ModelsJournal of Endodontics, 2000
- Defects in Rotary Nickel-Titanium Files After Clinical UseJournal of Endodontics, 2000
- Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instrumentsJournal of Endodontics, 1999
- Cyclic fatigue of Profile rotary instruments after simulated clinical useInternational Endodontic Journal, 1999
- Analysis of forces developed during mechanical preparation of extracted teeth using Profile NiTi rotary instrumentsInternational Endodontic Journal, 1999
- Cyclic fatigue testing of nickel-titanium endodontic instrumentsJournal of Endodontics, 1997
- A comparison of root canal preparations using Ni-Ti hand, Ni-Ti engine-driven, and K-Flex endodontic instrumentsJournal of Endodontics, 1995
- An initial investigation of the bending and torsional properties of nitinol root canal filesJournal of Endodontics, 1988