  • Accuracy of the TCM Endo III Torque-Control Motor for Nickel-Titanium Rotary Instruments

    by Ghassan Yared, Gajanan K. Kulkarni

    The purpose of this study was to determine the torque output and examine the accuracy of five TCM Endo III torque control motors. A handpiece was attached to the motor and gripped with a vise. A 0.07 taper Orifice Shaper, size 50, was inserted in the handpiece. The instrument tip was clamped in a chuck connected to a torque sensor. Four torque settings were evaluated at 350 rpm. Ten tests were performed at each torque setting. A new instrument was used for each test. The means of the torque values generated by the motors at the different torque levels were analyzed using analysis of variance and post-hoc pairwise comparisons with the Bonferroni test. The actual torque values were significantly higher than the torque preset on the motor (p < 0.0001) and did not differ significantly among the motors (p > 0.05). The actual torque deviated from the preset torque. The usefulness of these motors is questionable.


  • In Vitro Study of the Torsional Properties of New and Used ProFile Nickel Titanium Rotary Files

    by Ghassan Yared

    The purpose of this study was to investigate torque and angle of rotation at fracture of new and used .06 ProFile instruments. ProFile sizes 15 to 40 were tested according to American National Standards Institute/American Dental Association Specification #28. Thirty new and used instruments were tested for each size. An analysis of variance was used to compare the torque and angle of rotation at fracture between new and used instruments. The relationship between torque at fracture and size of instrument was determined with a regression analysis. The used instruments had significantly lower torque and angle of rotation at fracture values compared with the new instruments (p < 0.0001). A stronger relationship was found between the size of the instrument and the torque at fracture for the new instruments (p < 0.0001) than for the used ones (p < 0.0001). In conclusion, the repeated use of .06 ProFile instruments significantly reduced the torque and angle of rotation at fracture.


  • Torsional fatigue and endurance limit of a size 30 .06 ProFile rotary instrument

    by S. Best, P. Watson, R. Pilliar, G. G. K. Kulkarni, G. Yared

    Aim To evaluate the torsional cyclic fatigue characteristics and specifically the endurance limit (EL) of a nickel–titanium rotary instrument. Methodology Size 30 .06 taper ProFile instruments were evaluated. The equipment was assembled according to the ANSI/ADA Specification No. 28. The motor was programmed to repeatedly rotate to a selected deflection angle (DA) and then return to zero (cycle). Testing started at 200° and was continued at decreasing angles until 106 cycles were reached without instrument fracture. Ten instruments were tested at each DA. The mean log number of cycles to fracture and standard deviation were determined for each DA at which fracture occurred. The DA at which 106 cycles were reached without instrument fracture corresponded by definition to the EL. Analysis of variance and pairwise comparisons using Duncan's multiple range test were performed to detect significant differences among the mean log number of cycles of the different DA. Significance was determined at the 95% confidence level. Results Instruments cycled at larger DA consistently demonstrated fewer cycles to fracture than those cycled at smaller DA. The differences among the mean log number of cycles of the different DA were statistically significant (P < 0.001). Cycles of 106 were completed without instrument fracture at 2.5°. Conclusions A torsional fatigue profile was generated for a specific NiTi rotary instrument. The EL was 2.5°.


  • Accuracy of the DTC torque control motor for nickel‐titanium rotary instruments

    by G. Yared, G. K. Kulkarni

    Aim  To determine the torque output and examine the accuracy of five identical DTC torque control (DTC) motors.Methodology  Torque settings on the DTC motors for the .04 ISO ProFile nickel‐titanium (NiTi) rotary instruments were evaluated at 350 r.p.m. A modification of the ANSI/ADA Specification No. 28 setup for the evaluation of torsional properties of endodontic instruments was used. A handpiece was attached to the motor and gripped with a vice. An .07 Orifice Shaper, size 50, was inserted in the handpiece. The instrument tip was clamped in a chuck connected to a torque sensor. The motor was then activated to rotate the instrument in a clockwise direction until reversal of the rotation occurred. The actual torque generated at the reversal of the rotation was recorded. Ten tests were carried out at each torque setting. A new Orifice Shaper was used for each test. The means of the actual torque values generated by the motors at the different torque settings were compared with the torque values claimed by the manufacturer and analysed using analysis of variance and the Student's t‐test (α = 0.05).Results  The actual torque values were significantly higher than the torque preset on the motors (P < 0.0001) and did not differ significantly among the motors (P > 0.05).Conclusions  The actual torque deviated from the preset torque and was higher than the reported torque at fracture of several NiTi rotary instruments. The usefulness of these motors is questionable.


  • An in vitro study of the torsional properties of new and used K3 instruments

    by G. Yared, G. K. Kulkarni, F. Ghossayn

    Aim To compare torque and angle of rotation at fracture of new and used K3 rotary instruments. The relation between size of instrument and torque at fracture was also investigated. Methodology Nickel–titanium (Ni‐Ti) .06 K3 rotary instruments were used in a crown‐down manner at 300 r.p.m. to prepare canals in resin blocks. The torque and angle of rotation at fracture of new and the used Ni‐Ti .06 K3 rotary instruments (sizes 15–40) were determined according to ANSI/ADA Specification no. 28. Analysis of variance was used to compare the torque and angle of rotation at fracture amongst the different sizes of the new instruments and between new instruments and instruments of the same size, which had been used in resin blocks (α = 0.05). The relationship between torque at fracture and size of instrument was subjected to regression analysis. Results Torque at fracture of the new instruments increased significantly with the diameter. The used instruments (sizes 15, 20, 30, 35 and 40) had lower torque at fracture compared to the new ones (P < 0.0001). The means of angle of rotation at fracture between the different sizes of new instruments were significantly different (P < 0.0001) except for sizes 15–20 (P = 0.2561). The used instruments (sizes 20–40) had lower angle of rotation at fracture compared to the new ones (P < 0.05). A linear relationship was found between the size of the file and the torque at fracture for the new instruments (r2 = 0.84; P < 0.0001) and the used ones (r2 = 0.82; P < 0.0001). Conclusions In general, the results suggested that the torque and angle of rotation at fracture were significantly affected by the repeated use of .06 K3 instruments in resin blocks.


  • Torsional Properties Of New And Used Rotary K3 NiTi Files

    by Ghassan Yared, Gajanan K. Kulkarni, Fadia Ghossayn

    The purpose of this study was to compare torque (gcm) and angle of rotation (degrees) at fracture of new and used NiTi K3 .04 rotary instruments. The relation between size of instrument and torque at fracture was also investigated. The torque and angle of rotation at fracture of new and used NiTi K3 .04 rotary instruments sizes 15 to 40 were determined according to ANSI/ADA Specification No. 28. Analysis of variance was used to compare the torque and angle of rotation at fracture among the different sizes of the new instruments and between new and used instruments of the same size (α = 0.05). The relationship between torque at fracture and size of instrument was determined with a regression analysis. Torque at fracture of the new instruments increased significantly with the diameter (p < 0.05). The used instruments, sizes 25 to 40, had significantly lower torque at fracture values compared to the new ones (p < 0.05). The used instruments, sizes 20 and 35 had significantly lower angle of rotation at fracture compared to the new ones (p < 0.05). A stronger relationship was found between the size of the file and the torque at fracture for the new instruments (p < 0.0001) compared to the used ones (p < 0.0001). The results of the present study suggested that the torque at fracture values of new instruments increased significantly with the diameter. The results also suggested that repeated use of .04 K3 instruments affected mainly the torque at fracture.


  • Behaviour Of Hero NiTi Instruments Used By An Experienced Operator Under Access Limitations

    by G.M. Yared

    The purpose of this study was to evaluate the failure incidence of Hero 642 nickel‐titanium (NiTi) rotary instruments (HRI) when used by an experienced operator in conjunction with different motors and under limitations of access. HRI were used by an experienced operator in a crown‐down technique. The teeth were mounted in resin arches of a phantom head to reproduce limitations of access. In Groups 1, 2, 3 and 4, air, high torque control, low torque control and very low torque control motors were used respectively. Each group included 30 canals in extracted human molars. One set of HRI was used for each canal; the HRI were sterilised before each case. A 2.5% NaOCl solution was used as an irrigant. An experienced operator performed all the cleaning and shaping procedures after application of a rubber dam. The opening of the mouth in the phantom head was standardised for the four groups. Instrument fracture did not occur in any of the groups. The air, high torque control and low torque control motors were as safe as the very low torque control motor when HRI were used by an experienced operator in a crown‐down manner at 300 rpm.