Design and application of a gas-liquid separator for the removal of carbon dioxide in the eluent of an ion chromatography system
Authors:
Open Access
- 1 April 2025
- journal article
- abstracts
- Published by China Science Publishing & Media Ltd. in Chinese Journal of Chromatography
- Vol. 43 (4) , 382-387
Abstract
A gas-liquid separator (GLS) for the removal of carbon dioxide in the eluent of an ion chromatography (IC) system was developed. Firstly, the microporous hollow fiber polypropylene tube (PP-T) was inserted into the polytetrafluoroethylene tube (PTFE-T) with suitable inner diameter to form a double casing structure and wound into a spiral shape. Each terminal of the PP-T is 3 cm longer than that of the PTFE-T. Then the terminals of PP-T and PTFE-T were fixed respectively on two horizontal joints equipped with tee joints, and the regenerated liquid pipeline was fixed on a vertical arm interface equipped with tee joints. When the carbonate eluent flows into PP-T from the upstream suppressor, the inhibited carbonate solution will decompose and produce CO2, which then escapes through PP-T and enters the flowing absorption solution located in the annular space between PP-T and PTFE-T, so as to achieve continuous and effective CO2 removal. The production conditions and operating conditions of GLS were optimized, including the length of PP-T, the inner diameter of PTFE-T, the operating temperature, the type, concentration and flow rate of the absorption solution. The experimental results show that when 40 mmol/L potassium hydroxide solution is used as absorption solution, the removal efficiency of CO2 can reach more than 98% at the flow rate of 1 mL/min. Higher operating temperature is helpful to improve the removal efficiency of CO2 and reduce the baseline noise. The optimized operating temperature in this study was 40 ℃. When applying this GLS to the IC system for carbonate eluent, using a mixed eluent of 1.8 mmol/L potassium carbonate-3.2 mmol/L potassium bicarbonate (1∶1, v/v), the background conductance signal decreased from 41.6 mV without GLS to 5.5 mV after using GLS. Using a mixed standard solution of common anions (F-、Cl-、[Formula: see text] on the separation of other anions.Keywords
This publication has 3 references indexed in Scilit:
- A bipolar membrane-based cation electrolytic membrane suppressor for ion chromatographyJournal of Chromatography A, 2019
- Online Gas-Free Electrodialytic KOH Eluent Generator for Ion ChromatographyAnalytical Chemistry, 2018
- Asymmetric Membrane Fiber-Based Carbon Dioxide Removal Devices for Ion ChromatographyAnalytical Chemistry, 2004