Assessment of the image quality of virtual monochromatic spectral computed tomography images: a phantom study considering object contrast, radiation dose, and frequency characteristics
Authors:
- 5 January 2021
- journal article
- research article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 14 (1) , 41-49
Abstract
Fast kilovoltage (kVp)-switching technology cannot obtain conventional 120 kVp images; thus, 70 keV virtual monochromatic spectral computed tomography (CT) images (VMSI) are generally used. The contrast-to-noise ratio (CNR) is used to evaluate the image quality of VMSI; however, CNR does not include frequency characteristics. The present study aimed to investigate the evaluation methods of VMSI considering frequency characteristics by comparing the image quality of 70 keV VMSI with that of conventional 120 kVp images. The evaluated object contrasts were 70 and 300 Hounsfield units (HU). Scans used two radiation dose levels: low (LD) and standard (SD). The volume CT dose index of LD and SD was 4.8- and 12 mGy, respectively. Images were reconstructed by filtered back projection, evaluating CNR, noise power spectrum (NPS), task transfer function (TTF), and system performance (SP) function calculated as TTF2/ NPS. The total NPS values (spatial frequency range: 0.2 ~ 0.4 mm−1) of 70 keV VMSI were higher than those of 120 kVp images. The spatial frequency TTF values that reached 10% (f10%) of the 70 keV VMSI changed based on object contrast. For the low-contrast condition, a lower f10% was observed with 70 keV VMSI. The CNR of 70 keV VMSI was comparable to that of 120 kVp images in low- and high-contrast conditions. However, for 70 keV VMSI, SP of low-contrast was low, and SP of high-contrast was high, compared with those of 120 kVp images. This study suggested that only CNR was not sufficient to evaluate the image quality of VMSI; thus, evaluation methods considering frequency characteristics should be used.Keywords
This publication has 28 references indexed in Scilit:
- High-Pitch Wide-Coverage Fast-Kilovoltage-Switching Dual-Energy CT: Impact of Pitch on Noise, Spatial Resolution, and Iodine Quantification in a Phantom StudyAmerican Journal of Roentgenology, 2019
- Virtual non-enhanced dual-energy CT reconstruction may replace true non-enhanced CT scans in the setting of suspected active hemorrhageEuropean Journal of Radiology, 2018
- Image quality characteristics for virtual monoenergetic images using dual-layer spectral detector CT: Comparison with conventional tube-voltage imagesPhysica Medica, 2018
- Development of a dual-energy computed tomography quality control program: Characterization of scanner response and definition of relevant parameters for a fast-kVp switching dual-energy computed tomography systemMedical Physics, 2018
- Fast kVp‐switching dual energy contrast‐enhanced thorax and cardiac CT: A phantom study on the accuracy of iodine concentration and effective atomic number measurementMedical Physics, 2017
- Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CTEuropean Radiology, 2016
- Performance of today's dual energy CT and future multi energy CT in virtual non‐contrast imaging and in iodine quantification: A simulation studyMedical Physics, 2015
- Hypodense liver lesions in patients with hepatic steatosis: do we profit from dual-energy computed tomography?European Radiology, 2015
- Abdominal CT: An intra-individual comparison between virtual monochromatic spectral and polychromatic 120-kVp images obtained during the same examinationEuropean Journal of Radiology, 2014
- Dual-Energy CT Angiography for Assessment of Regional Pulmonary Perfusion in Patients With Chronic Thromboembolic Pulmonary Hypertension: Initial ExperienceAmerican Journal of Roentgenology, 2011