A radiotherapy community data‐driven approach to determine which complexity metrics best predict the impact of atypical TPS beam modeling on clinical dose calculation accuracy
Open Access
- 1 March 2024
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (5) , e14318
Abstract
Purpose: To quantify the impact of treatment planning system beam model parameters, based on the actual spread in radiotherapy community data, on clinical treatment plans and determine which complexity metrics best describe the impact beam modeling errors have on dose accuracy.Methods: Ten beam modeling parameters for a Varian accelerator were modified in RayStation to match radiotherapy community data at the 2.5, 25, 50, 75, and 97.5 percentile levels. These modifications were evaluated on 25 patient cases, including prostate, non‐small cell lung, H&N, brain, and mesothelioma, generating 1,000 plan perturbations. Differences in the mean planned dose to clinical target volumes (CTV) and organs at risk (OAR) were evaluated with respect to the planned dose using the reference (50th‐percentile) parameter values. Correlation between CTV dose differences, and 18 different complexity metrics were evaluated using linear regression; R‐squared values were used to determine the best metric.Results: Perturbations to MLC offset and transmission parameters demonstrated the greatest changes in dose: up to 5.7% in CTVs and 16.7% for OARs. More complex clinical plans showed greater dose perturbation with atypical beam model parameters. The mean MLC Gap and Tongue & Groove index (TGi) complexity metrics best described the impact of TPS beam modeling variations on clinical dose delivery across all anatomical sites; similar, though not identical, trends between complexity and dose perturbation were observed among all sites.Conclusion: Extreme values for MLC offset and MLC transmission beam modeling parameters were found to most substantially impact the dose distribution of clinical plans and careful attention should be given to these beam modeling parameters. The mean MLC Gap and TGi complexity metrics were best suited to identifying clinical plans most sensitive to beam modeling errors; this could help provide focus for clinical QA in identifying unacceptable plans.Keywords
Funding Information
- National Institutes of Health
- National Cancer Institute (CA180803, CA214526)
This publication has 34 references indexed in Scilit:
- Dosimetric and mechanical equivalency of Varian TrueBeam linear acceleratorsJournal of Applied Clinical Medical Physics, 2020
- Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven dataMedical Physics, 2020
- Impact of the MLC leaf‐tip model in a commercial TPS: Dose calculation limitations and IROC‐H phantom failuresJournal of Applied Clinical Medical Physics, 2020
- Reference dataset of users’ photon beam modeling parameters for the Eclipse, Pinnacle, and RayStation treatment planning systemsMedical Physics, 2019
- On the need for tuning the dosimetric leaf gap for stereotactic treatment plans in the Eclipse treatment planning systemJournal of Applied Clinical Medical Physics, 2019
- Comparison of complexity metrics for multi-institutional evaluations of treatment plans in radiotherapyPhysics and Imaging in Radiation Oncology, 2018
- Relationship between dosimetric leaf gap and dose calculation errors for high definition multi-leaf collimators in radiotherapyPhysics and Imaging in Radiation Oncology, 2018
- Treatment Planning System Calculation Errors Are Present in Most Imaging and Radiation Oncology Core-Houston Phantom FailuresInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Agreement Between Institutional Measurements and Treatment Planning System Calculations for Basic Dosimetric Parameters as Measured by the Imaging and Radiation Oncology Core-HoustonInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2016
- Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapyRadiotherapy and Oncology, 2010