On the need for tuning the dosimetric leaf gap for stereotactic treatment plans in the Eclipse treatment planning system
Open Access
- 21 June 2019
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 20 (7) , 68-77
Abstract
The dosimetric leaf gap (DLG) and tongue‐and‐groove (T&G) effects are critical aspects in the modeling of multileaf collimators (MLC) in the treatment planning system (TPS). In this study, we investigated the dosimetric impact of limitations associated with the T&G modeling in stereotactic plans and its relationship with the need for tuning the DLG in the Eclipse TPS. Measurements were carried out using Varian TrueBeam STx systems from two different institutions. Test fields presenting MLC patterns with several MLC gap sizes (meanGap) and different amounts of T&G effect (TGi) were first evaluated. Secondly, dynamic conformal arc (DCA) and volumetric modulated arc therapy (VMAT) deliveries of stereotactic cases were analyzed in terms of meanGap and TGi. Two DLG values were used in the TPS: the measured DLG (DLGmeas) and an optimal DLG (DLGopt). Measured and calculated doses were compared according to dose differences and gamma passing rates (GPR) with strict local gamma criteria of 2%/2 mm. The discrepancies were analyzed for DLGmeas and DLGopt, and their relationships with both TGi and meanGap were investigated. DCA arcs involved significantly lower TGi and larger meanGap than VMAT arcs (P < 0.0001). By using DLGmeas in the TPS, the dose discrepancies increased as TGi increased and meanGap decreased for both test fields and clinical plans. Dose discrepancies dramatically increased with the ratio TGi/meanGap. Adjusting the DLG value was then required to achieve acceptable calculations and configuring the TPS with DLGopt led to an excellent agreement with median GPRs (2%/2 mm) > 99% for both institutions. We also showed that DLGopt could be obtained from the results of the test fields. We demonstrated that the need for tuning the DLG is due to the limitations of the T&G modeling in the Eclipse TPS. A set of sweeping gap tests modified to incorporate T&G effects can be used to determine the optimal DLG value.Keywords
This publication has 35 references indexed in Scilit:
- Verification of dosimetric accuracy on the TrueBeam STx: Rounded leaf effect of the high definition MLCMedical Physics, 2012
- Leaf Gap Leakage and Dosimetric Accuracy for SBRT Treatment on the TrueBeam STxInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Implementing RapidArc into clinical routine: A comprehensive program from machine QA to TPS validation and patient QAMedical Physics, 2011
- On the use of the MLC dosimetric leaf gap as a quality control tool for accurate dynamic IMRT deliveryMedical Physics, 2011
- Stereotactic radiotherapy for peripheral lung tumors: A comparison of volumetric modulated arc therapy with 3 other delivery techniquesRadiotherapy and Oncology, 2010
- An oscillating sweeping gap test for VMAT quality assurancePhysics in Medicine & Biology, 2010
- IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119Medical Physics, 2009
- A generalized a priori dose uncertainty model of IMRT deliveryMedical Physics, 2008
- An independent dose calculation algorithm for MLC-based radiotherapy including the spatial dependence of MLC transmissionPhysics in Medicine & Biology, 2008
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998