Validation of the thermal stability of EMAIS-SR autonomous packaging for preservation of porcine lungs: cooling curve in an experimental model
Authors:
- Flavio Pola dos ReisFRFlavio Pola dos Reis
- Pedro Henrique Andrade GuimarãesPGPedro Henrique Andrade Guimarães
- Matheus Werner FantatoMFMatheus Werner Fantato
- Selma Ferrer da SilvaSSSelma Ferrer da Silva
- Liliane Moreira RuizLRLiliane Moreira Ruiz
- Natalia Aparecida NepomucenoNNNatalia Aparecida Nepomuceno
- Edson Caoru KitaniEKEdson Caoru Kitani
- Bartira de Aguiar RozaBRBartira de Aguiar Roza
- Paulo Manuel Pêgo FernandesPFPaulo Manuel Pêgo Fernandes
Open Access
- 1 January 2026
- journal article
- Published by FapUNIFESP (SciELO) in Sao Paulo Medical Journal
- Vol. 144 (4)
Abstract
BACKGROUND: Intermediate temperatures (10°C) better preserve mitochondrial function and extend ischemic time, which are fundamental for lung preservation during transplantation. The EMAIS-SR device was developed as a reusable, cost-effective alternative to maintain the 10°C set-point, thereby overcoming logistical and economic barriers in organ transport. OBJECTIVE: This study aimed to validate the thermal stability of the EMAIS-SR system in a swine model by evaluating the organ-cooling curve and the time spent within the target temperature range. DESIGN AND SETTING: Validation study using experimental surgery on porcine organs. METHODS: Lung blocks were harvested from two Landrace swine in independent experiments (P1, P2). The organs were placed in triple sterile packaging. Thermal stability was monitored for 48 h by using four negative temperature coefficient (NTC) thermistors (three intrapulmonary and one extrapulmonary) with data acquisition at a 0.1-s interval. Another datalogger (Hantek ® HTM208B) was used for control and validation. Thermal distribution was further characterized by infrared thermography (FLIR C3-X). RESULTS: In Experiment 1 (P1), intrapulmonary temperatures reached the predefined target range (≤ 12°C) approximately 1.4 h after lung extraction, and the mean temperature recorded in the container was 8.01 °C (standard deviation, 2.88°C). In Experiment 2 (P2), a mean temperature reduction of 6.6°C was observed, bringing the intrapulmonary temperature to the target range (≤ 12°C) within approximately 2 h post-extraction. CONCLUSION: The EMAIS-SR device effectively maintained lung-preservation temperatures within the target 8°C–12°C range over 48 h. These results demonstrate the system’s technical feasibility and its capacity to provide consistent thermal stability across intrapulmonary compartments.Keywords
This publication has 13 references indexed in Scilit:
- Short-term outcomes after lung transplantation using grafts preserved at 10 °c: A single-center studyEuropean Journal of Cardio-Thoracic Surgery, 2025
- Lung transplant outcomes after implementation of a hospital-based 10 °C controlled hypothermic organ preservation unitThe Journal of Thoracic and Cardiovascular Surgery, 2025
- Temperature dynamics of donor lungs from procurement to reperfusion: Static ice versus controlled hypothermic storageThe Journal of Heart and Lung Transplantation, 2025
- Prolonged Preservation of up to 24 Hours at 10 °C Does Not Impact Outcomes after Lung TransplantationAnnals of Surgery, 2025
- The 2024 American Association for Thoracic Surgery expert consensus document: Current standards in donor lung procurement and preservationThe Journal of Thoracic and Cardiovascular Surgery, 2025
- The effect of rewarming ischemia on tissue transcriptome and metabolome signatures: A clinical observational study in lung transplantationThe Journal of Heart and Lung Transplantation, 2024
- Protective effects of 10 °C preservation on donor lungs with lipopolysaccharide-induced acute lung injuryThe Journal of Thoracic and Cardiovascular Surgery, 2024
- Evaluation of 10°C as the optimal storage temperature for aspiration-injured donor lungs in a large animal transplant modelThe Journal of Heart and Lung Transplantation, 2022
- A method for safe twelve-hour pulmonary preservationThe Journal of Thoracic and Cardiovascular Surgery, 1989
- The effect of ischemic time and temperature on lung preservation in a simple ex vivo rabbit model used for functional assessmentThe Journal of Thoracic and Cardiovascular Surgery, 1989