New analytical approach to calibrate the NaI (Tl) detectors using spherical radioactive sources
- 17 March 2013
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 156 (1) , 109-117
Abstract
A new theoretical approach was used to calibrate and calculate the full-energy peak efficiency of the NaI (Tl) detectors based on the direct statistical method proposed by Selim and Abbas for cylindrical detectors. In addition, the self-attenuation of the source matrix, the attenuation by the source container and the detector housing materials were considered in the mathematical treatment. Results were compared with those measured by a cylindrical NaI (Tl) detector with resolution (FWHM) at 662 keV equal to 7.5 %. 152Eu aqueous radioactive spherical sources covering the energy range from 121 to 1408 keV were used. In comparison, the calculated and the measured full-energy peak efficiency values were in good agreement.Keywords
This publication has 15 references indexed in Scilit:
- Calculation of relative full-energy peak efficiencies of well-type detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- Analytical formulae for well-type NaI (Tl) and HPGe detectors efficiency computationApplied Radiation and Isotopes, 2001
- A direct mathematical method to calculate the efficiencies of a parallelepiped detector for an arbitrarily positioned point sourceRadiation Physics and Chemistry, 2001
- Analytical calculation of the efficiencies of gamma scintillators. Part I: Total efficiency for coaxial disk sourcesRadiation Physics and Chemistry, 1998
- HPGe detector efficiency calibration for extended cylinder and marinelli-beaker sources using the ESOLAN programApplied Radiation and Isotopes, 1997
- Self-attenuation and coincidence-summing corrections calculated by Monte Carlo simulations for gamma-spectrometric measurements with well-type germanium detectorsApplied Radiation and Isotopes, 1996
- HPGe detector absolute-peak-efficiency calibration by using the ESOLAN programApplied Radiation and Isotopes, 1995
- Application of a new Monte Carlo method for determination of summation and self-attenuation corrections in gamma spectrometryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993
- Detector-efficiency calculation based on point-source measurementThe International Journal of Applied Radiation and Isotopes, 1983
- Calculation of the peak efficiency of high-purity germanium detectorsThe International Journal of Applied Radiation and Isotopes, 1983