Repeatability of the ciliary sulcus-to-sulcus diameter measurement using wide-scanning-field ultrasound biomicroscopy
Authors:
- Takashi KojimaTKTakashi Kojima
- Rie HoraiRHRie Horai
- Mayuka ItoMIMayuka Ito
- Tomoaki NakamuraTNTomoaki Nakamura
- Kazuo IchikawaKIKazuo Ichikawa
- 1 July 2011
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 37 (7) , 1251-1256
Abstract
To evaluate the interexaminer and intraexaminer variance in ciliary sulcus-to-sulcus (STS) diameter measurements using wide-scanning-field ultrasound biomicroscopy (UBM) (Vumax II). Nagoya Eye Clinic, Nagoya, Japan. Evaluation of diagnostic test or technology. Two examiners measured the STS diameter 10 times in each subject. Intraexaminer and interexaminer variances were calculated. To determine interexaminer variance at each UBM measurement step, the examiner was changed every 2 steps when the STS diameter was measured. The Bland-Altman plot test was used to analyze each step of measurement variance. Thirty eyes of 15 volunteers (9 men, 6 women; mean age 29.6 years ± 6.2 [SD]) were enrolled. The mean coefficient of variation was 0.62% ± 0.20% for intraexaminer measurements and 3.4% ± 2.4% for interexaminer measurements. There were statistically significant differences between the 2 examiners in the step of extracting images and determining and measuring the STS diameter (P=.02). The 95% confidence intervals in all steps were relatively high (recordings, -1.02 to 1.08 mm; extracting images and determining and measuring STS; -0.79 to 1.25 mm). Interexaminer variance in STS measurement should be considered if the STS diameter is used for selecting the size of posterior chamber phakic intraocular lenses.Keywords
This publication has 19 references indexed in Scilit:
- Relationship between ciliary sulcus diameter and anterior chamber diameter and corneal diameterJournal of Cataract & Refractive Surgery, 2010
- Correlation between ciliary sulcus diameter measured by 35 MHz ultrasound biomicroscopy and other ocular measurementsJournal of Cataract & Refractive Surgery, 2008
- Direct Measurement of the Ciliary Sulcus Diameter by 35-Megahertz Ultrasound BiomicroscopyOphthalmology, 2007
- Magnetic resonance imaging and Orbscan assessment of the anterior chamberJournal of Cataract & Refractive Surgery, 2005
- Correlation between different measurements within the eye relative to phakic intraocular lens implantationJournal of Cataract & Refractive Surgery, 2004
- Comparison of manual and automated methods to determine horizontal corneal diameterJournal of Cataract & Refractive Surgery, 2004
- Implantable contact lens for moderate to high myopiaJournal of Cataract & Refractive Surgery, 2003
- U.S. Food and Drug Administration clinical trial of the Implantable Contact Lens for moderate to high myopiaOphthalmology, 2003
- Artisan phakic intraocular lens for myopia Short-term results of a prospective, multicenter studyOphthalmology, 2002
- Cataract development after implantationof the Staar Collamer posterior chamber phakic lensJournal of Cataract & Refractive Surgery, 1999