An evaluation of exhaled flow measuring mouthpieces for breath sampling devices
Authors:
- 1 August 2014
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
- Vol. 2014 (1094687X) , 5768-5771
Abstract
A study is conducted to determine the dimensions of an exhaled flow measuring mouthpiece for a breath-sampling device that requires breathing for an extended period of time. Fleisch, Pitot and Venturi type differential pressure based flow measuring mouthpieces of various dimensions are evaluated. Inner diameters (IDs) of the cylinder shaped Fleisch, Pitot and Venturi mouthpieces are varied from 5mm to 25mm. Based on the study, we conclude IDs ranging from 8.75mm to 12.5mm for Fleisch type, IDs ranging from 10mm to 12.5mm for Pitot type and ID of 10mm for Venturi type are the most suitable dimensions for a breath sampling device.Keywords