Evaluation of the RF heating of a generic deep brain stimulator exposed in 1.5 T magnetic resonance scanners
Authors:
- 11 October 2012
- journal article
- research article
- Published by Wiley in Bioelectromagnetics
- Vol. 34 (2) , 104-113
Abstract
The radio frequency (RF) electromagnetic field of magnetic resonance (MR) scanners can result in significant tissue heating due to the RF coupling with the conducting parts of medical implants. The objective of this article is to evaluate the advantages and shortcomings of a new four‐tier approach based on a combined numerical and experimental procedure, designed to demonstrate safety of implants during MR scans. To the authors' best knowledge, this is the first study analyzing this technique. The evaluation is performed for 1.5 T MR scanners using a generic model of a deep brain stimulator (DBS) with a straight lead and a helical lead. The results show that the approach is technically feasible and provides sound and conservative information about the potential heating of implants. We demonstrate that (1) applying optimized tools results in reasonable uncertainties for the overall evaluation; (2) each tier reduces the overestimation by several dB at the cost of more demanding evaluation steps; (3) the implant with the straight lead would cause local temperature increases larger than 18 °C at the RF exposure limit for the normal operating mode; (4) Tier 3 is not sufficient for the helical implant; and (5) Tier 4 might be too demanding to be performed for complex implants. We conclude with a suggestion for a procedure that follows the same concept but is between Tier 3 and 4. In addition, the evaluation of Tier 3 has shown consistency with current scan practice, namely, the resulting heat at the lead tip is less than 3.5 °C for the straight lead and 0.7 °C for the helix lead for scans at the current applied MR scan restrictions for deep brain stimulation at a head average SAR of 0.1 W/kg. Bioelectromagnetics 34:104–113, 2013.Keywords
This publication has 20 references indexed in Scilit:
- Local tissue temperature increase of a generic implant compared to the basic restrictions defined in safety guidelinesBioelectromagnetics, 2011
- Analysis of the Role of Lead Resistivity in Specific Absorption Rate for Deep Brain Stimulator Leads at 3T MRIIEEE Transactions on Medical Imaging, 2010
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- Measurement, simulation and uncertainty assessment of implant heating during MRIPhysics in Medicine & Biology, 2009
- Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurationsBioMedical Engineering OnLine, 2008
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heatingJournal of Magnetic Resonance Imaging, 2004
- Magnetic resonance imaging and cardiac pacemaker safety at 1.5-TeslaJACC, 2004
- The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissuesPhysics in Medicine & Biology, 1996
- Insulated linear antenna: Theory and experimentJournal of Applied Physics, 1974