Influence of the Acetabular Cup Material on the Shell Deformation and Strain Distribution in the Adjacent Bone—A Finite Element Analysis
Open Access
Abstract
In total hip arthroplasty, excessive acetabular cup deformations and altered strain distribution in the adjacent bone are potential risk factors for implant loosening. Materials with reduced stiffness might alter the strain distribution less, whereas shell and liner deformations might increase. The purpose of our current computational study was to evaluate whether carbon fiber-reinforced poly-ether-ether-ketones with a Young´s modulus of 15 GPa (CFR-PEEK-15) and 23 GPa (CFR-PEEK-23) might be an alternative shell material compared to titanium in terms of shell and liner deformation, as well as strain distribution in the adjacent bone. Using a finite element analysis, the press-fit implantation of modular acetabular cups with shells made of titanium, CFR-PEEK-15 and CFR-PEEK-23 in a human hemi-pelvis model was simulated. Liners made of ceramic and polyethylene were simulated. Radial shell and liner deformations as well as strain distributions were analyzed. The shells made of CFR-PEEK-15 were deformed most (266.7 µm), followed by CFR-PEEK-23 (136.5 µm) and titanium (54.0 µm). Subsequently, the ceramic liners were radially deformed by up to 4.4 µm and the polyethylene liners up to 184.7 µm. The shell materials slightly influenced the strain distribution in the adjacent bone with CFR-PEEK, resulting in less strain in critical regions (3000 µm/m) and more strain in bone building or sustaining regions (400 to 3000 µm/m), while the liner material only had a minor impact. The superior biomechanical properties of the acetabular shells made of CFR-PEEK could not be determined in our present study.Keywords
This publication has 62 references indexed in Scilit:
- In vivo implant fixation of carbon fiber‐reinforced PEEK hip prostheses in an ovine modelJournal of Orthopaedic Research, 2012
- High Rate of Ceramic Sandwich Liner FractureClinical Orthopaedics and Related Research, 2012
- Nano-TiO2/PEEK bioactive composite as a bone substitute material: in vitro and in vivo studiesInternational Journal of Nanomedicine, 2012
- The influence of acetabular cup material on pelvis cortex surface strains, measured using digital image correlationJournal of Biomechanics, 2012
- Factors Affecting Modular Acetabular Ceramic Liner Insertion: A Biomechanical AnalysisClinical Orthopaedics and Related Research, 2012
- Fractures of a sandwich ceramic liner at ten year follow-upInternational Orthopaedics, 2011
- Deformation of metal-backed acetabular components and the impact of liner thickness in a cadaveric modelInternational Orthopaedics, 2010
- Seating of Ceramic Liners in the Uncemented Trident® Acetabular Shell: Is There Really a Problem?Clinical Orthopaedics and Related Research, 2009
- Retroacetabular Stress-shielding in THAClinical Orthopaedics and Related Research, 2008
- Safety factors in bone strengthCalcified Tissue International, 1993