Effective magnetic susceptibility of 3D‐printed porous metal scaffolds
- 25 January 2022
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 87 (6) , 2947-2956
Abstract
Purpose: 3D‐printed porous metal scaffolds are a promising emerging technology in orthopedic implant design. Compared to solid metal implants, porous metal implants have lower magnetic susceptibility values, which have a direct impact on imaging time and image quality. The purpose of this study is to determine the relationship between porosity and effective susceptibility through quantitative estimates informed by comparing coregistered scanned and simulated field maps.Methods: Five porous scaffold cylinders were designed and 3D‐printed in titanium alloy (Ti‐6Al‐4V) with nominal porosities ranging from 60% to 90% using a cellular sheet‐based gyroid design. The effective susceptibility of each cylinder was estimated by comparing acquired B0 field maps against simulations of a solid cylinder of varying assigned magnetic susceptibility, where the orientation and volume of interest of the simulations was informed by a custom alignment phantom.Results: Magnitude images and field maps showed obvious decreases in artifact size and field inhomogeneity with increasing porosity. The effective susceptibility was found to be linearly correlated with porosity (R2 = 0.9993). The extrapolated 100% porous (no metal) magnetic susceptibility was −9.9 ppm, closely matching the expected value of pure water (−9 ppm), indicating a reliable estimation of susceptibility.Conclusion: Effective susceptibility of porous metal scaffolds is linearly correlated with porosity. Highly porous implants have sufficiently low effective susceptibilities to be more amenable to routine imaging with MRI.Keywords
Funding Information
- Canadian Institutes of Health Research (FDN 148474, PJT 159665)
This publication has 31 references indexed in Scilit:
- Characterizing the limits of MRI near metallic prosthesesMagnetic Resonance in Medicine, 2014
- Alias subtraction more efficient than conventional zero‐padding in the Fourier‐based calculation of the susceptibility induced perturbation of the magnetic field in MRMagnetic Resonance in Medicine, 2012
- Reflections concerning triply-periodic minimal surfacesInterface Focus, 2012
- Intervention‐based multidimensional phase unwrapping using recursive orthogonal referringMagnetic Resonance in Medicine, 2012
- Slice encoding for metal artifact correction with noise reductionMagnetic Resonance in Medicine, 2011
- Imaging near metal with a MAVRIC‐SEMAC hybridMagnetic Resonance in Medicine, 2010
- Signal decay due to susceptibility-induced intravoxel dephasing on multiple air-filled cylinders: MRI simulations and experimentsMagnetic Resonance Materials in Physics, Biology and Medicine, 2008
- Magnetic susceptibility quantification for arbitrarily shaped objects in inhomogeneous fieldsMagnetic Resonance in Medicine, 2001
- The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kindsMedical Physics, 1996
- Acceptance testing of magnetic resonance imaging systems: Report of AAPM Nuclear Magnetic Resonance Task Group No. 6Medical Physics, 1992