Slice encoding for metal artifact correction with noise reduction
Authors:
Open Access
- 1 February 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 65 (5) , 1352-1357
Abstract
Magnetic resonance imaging (MRI) near metallic implants is often hampered by severe metal artifacts. To obtain distortion‐free MR images near metallic implants, SEMAC (Slice Encoding for Metal Artifact Correction) corrects metal artifacts via robust encoding of excited slices against metal‐induced field inhomogeneities, followed by combining the data resolved from multiple SEMAC‐encoded slices. However, as many of the resolved data elements only contain noise, SEMAC‐corrected images can suffer from relatively low signal‐to‐noise ratio. Improving the signal‐to‐noise ratio of SEMAC‐corrected images is essential to enable SEMAC in routine clinical studies. In this work, a new reconstruction procedure is proposed to reduce noise in SEMAC‐corrected images. A singular value decomposition denoising step is first applied to suppress quadrature noise in multi‐coil SEMAC‐encoded slices. Subsequently, the singular value decomposition‐denoised data are selectively included in the correction of through‐plane distortions. The experimental results demonstrate that the proposed reconstruction procedure significantly improves the SNR without compromising the correction of metal artifacts. Magn Reson Med, 2011.Keywords
Funding Information
- NTU Start-up Grant, MOE AcRF Tier 1 RG 25/08 (NIH-R21-EB008190, NIH-P41-RR009784)
This publication has 15 references indexed in Scilit:
- Accelerated slice encoding for metal artifact correctionJournal of Magnetic Resonance Imaging, 2010
- SEMAC: Slice encoding for metal artifact correction in MRIMagnetic Resonance in Medicine, 2009
- A multispectral three‐dimensional acquisition technique for imaging near metal implantsMagnetic Resonance in Medicine, 2009
- Comprehensive quantification of signal‐to‐noise ratio and g‐factor for image‐based and k‐space‐based parallel imaging reconstructionsMagnetic Resonance in Medicine, 2008
- Reduction of blurring in view angle tilting MRIMagnetic Resonance in Medicine, 2005
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- De-noising by soft-thresholdingIEEE Transactions on Information Theory, 1995
- SVD-based quantification of magnetic resonance signalsJournal of Magnetic Resonance (1969), 1992
- Homodyne detection in magnetic resonance imagingIEEE Transactions on Medical Imaging, 1991