Stabilised Biosensing Using Needle‐Based Recess Electrodes
- 20 December 2011
- journal article
- conference paper
- Published by Wiley in Electroanalysis
- Vol. 24 (3) , 529-538
Abstract
A recess disk electrode for amperometric monitoring of oxygen and glucose is reported. The basic design of sensor is a needle structure for easy implantation, embodying a gold filled silica capillary with a 80 µm inner diameter working electrode. Recess and inlaid disc electrodes, placed inside stainless steel tubes which served as a counter/reference electrode, were compared. The working electrode surface was modified with different treatments and barrier membranes to achieve selectivity for the analytes of interest. The basic needle format is ideal for in vivo use, and the format is easily extendable to other analyte targets. The sensors show linear working range for oxygen up to 160 mmHg partial pressure of oxygen, and for glucose from 1 to 10 mM. Bio‐fouling, as assessed by exposure to bovine serum albumin, was significantly reduced. Response times for the recess construct was increased but remained within the acceptable range for physiological monitoring. The operational stability of the sensors is demonstrated as well as the interference‐free detection of peroxide in the presence of physiologically relevant levels of ascorbic acid, uric acid, acetaminophen, and catechol. Preliminary in vivo tests showed excellent response towards glucose.Keywords
This publication has 39 references indexed in Scilit:
- Thick-Film Carbon Dioxide Sensor via Anodic Adsorbate Stripping Technique and Its Structural DependenceSensors, 2009
- Development of a self‐cleaning sensor membrane for implantable biosensorsJournal of Biomedical Materials Research Part A, 2008
- A Bipartite Expression for the Transient Amperometric Current at a Membrane Covered Planar Electrode to Characterize Solute Diffusion Through the MembraneElectroanalysis, 2006
- Home blood glucose biosensors: a commercial perspectiveBiosensors and Bioelectronics, 2005
- A glucose-sensing contact lens: from bench top to patientCurrent Opinion in Biotechnology, 2005
- Disposable screen-printed biosensor for transcutaneous oxygen measurementMeasurement Science and Technology, 2002
- Diffusion restricting outer membranes for greatly extended linearity measurements with glucose oxidase enzyme electrodesAnalytica Chimica Acta, 1996
- Direct Enzyme-Amplified Electrical Recognition of a 30-Base Model OligonucleotideJournal of the American Chemical Society, 1996
- Stabilized needle electrode system for in vivo glucose monitoring based on open flow microperfusionThe Analyst, 1996
- Glucose enzyme electrode with extended linearity: Application to undiluted blood measurementsAnalytica Chimica Acta, 1986