Assessing effects of PEEP and global expiratory lung volume on regional electrical impedance tomography
Authors:
- D. G. MarkhorstDMD. G. Markhorst
- A. B. J. GroeneveldAGA. B. J. Groeneveld
- R. M. HeethaarRHR. M. Heethaar
- E. ZonneveldEZE. Zonneveld
- H. R. Van GenderingenHGH. R. Van Genderingen
- 1 January 2009
- journal article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 33 (4) , 281-287
Abstract
This study was performed to assess the value of electrical impedance tomography (EIT) as an indicator of tidal (V(T)) and end expiratory lung volume (EELV). EIT measurements were performed in seven healthy piglets during constant tidal volume ventilation at incremental and decremental positive end-expiratory pressure (PEEP) levels. Tidal impedance changes were calibrated to volume using V(T) calculated from flow at the airway opening. Simultaneously, calibrated respiratory inductive plethysmography was used to measure EELV changes, and used as a reference standard. EIT systematically underestimated both V(T) and EELV changes when EELV deviated from the level at which it was calibrated. Calculated over the entire pressure-volume curve, EIT systematically underestimated V(T) by 28 ml, with a precision from -16 to 72 ml. EELV was systemically underestimated by 406 ml, with a precision of -38 to 849 ml. Nonlinear recruitment in the ventral regions of the lungs was the main cause of this underestimation. Tidal and end-expiratory changes in pulmonary impedance reflect corresponding changes in lung volume, but the increasing underestimation with increasing lung volume should be taken into account in the analysis of EIT data.Keywords
This publication has 15 references indexed in Scilit:
- Respiratory inductive plethysmography accuracy at varying PEEP levels and degrees of acute lung injuryJournal of Medical Engineering & Technology, 2006
- Imbalances in Regional Lung VentilationAmerican Journal of Respiratory and Critical Care Medicine, 2004
- Regional Ventilation by Electrical Impedance TomographyChest, 2003
- End-expiratory lung impedance change enables bedside monitoring of end-expiratory lung volume changeIntensive Care Medicine, 2003
- Detection of local lung air content by electrical impedance tomography compared with electron beam CTJournal of Applied Physiology, 2002
- Regional distribution of gas and tissue in acute respiratory distress syndrome. III. Consequences for the effects of positive end-expiratory pressureIntensive Care Medicine, 2000
- A Scanographic Assessment of Pulmonary Morphology in Acute Lung Injury: Significance of the Lower Inflection Point Detected on the Lung Pressure–Volume CurveAmerican Journal of Respiratory and Critical Care Medicine, 1999
- Beneficial Effects of the "Open Lung Approach" With Low Distending Pressures in Acute Respiratory Distress Syndrome. A Prospective Randomized Study on Mechanical VentilationAmerican Journal of Respiratory and Critical Care Medicine, 1995
- A review of image reconstruction techniques for electrical impedance tomographyMedical Physics, 1989