Numerical evaluation of image homogeneity, signal-to-noise ratio, and specific absorption rate for human brain imaging at 1.5, 3, 7, 10.5, and 14T in an 8-channel transmit/receive array
Open Access
- 27 June 2014
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 41 (5) , 1432-1439
Abstract
Purpose To predict signal-to-noise ratio (SNR) trends and absorbed energy in magnetic resonance imaging (MRI) of the brain up to 14T. Materials and Methods A human head in an eight-channel transmit/receive coil was simulated with Maxwell and Bloch equations to determine excitation homogeneity with radiofrequency (RF) shimming, image homogeneity, SNR, and absorbed energy in MRI from 1.5 to 14T considering realistic field distributions and relaxation properties. Results RF shimming alone achieved a standard deviation in excitation flip angle less than 10° in mid-brain up to 14T, but produced a small region with low excitation on a lower slice. Current reconstruction methods may produce shading artifacts at 14T. SNR increases with a greater-than-linear rate for gradient recalled echo (GRE) sequences having short (2 msec) echo time (TE) and long relaxation time (TR) (∼2.3-fold increase from 7T to 14T), but a less-than-linear rate if TE is 10 msec (∼1.6-fold increase from 7T to 14T). Depending on the sequence, SNR per square root of imaging time may produce a less-than-linear increase with B0. Whole-head absorbed energy shows a less-than-quadratic increase with B0 (1.7-fold increase from 7T to 14T). Conclusion Numerical simulations indicate that with proper preparation and precautions, imaging of the human brain at up to 14T could be performed safely, with advantages in SNR. J. Magn. Reson. Imaging 2015;41:1432–1439.Keywords
This publication has 25 references indexed in Scilit:
- Calculation of radiofrequency electromagnetic fields and their effects in MRI of human subjectsMagnetic Resonance in Medicine, 2011
- T2* measurements in human brain at 1.5, 3 and 7 TMagnetic Resonance Imaging, 2007
- Spin‐echo fMRI in humans using high spatial resolutions and high magnetic fieldsMagnetic Resonance in Medicine, 2003
- 7T vs. 4T: RF power, homogeneity, and signal‐to‐noise comparison in head imagesMagnetic Resonance in Medicine, 2001
- Calculations of B1 distribution, SNR, and SAR for a surface coil adjacent to an anatomically‐accurate human body modelMagnetic Resonance in Medicine, 2001
- Signal‐to‐noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coilMagnetic Resonance in Medicine, 2001
- Sensitivity and Power Deposition in a High-Field Imaging ExperimentJournal of Magnetic Resonance Imaging, 2000
- Ultimate intrinsic signal‐to‐noise ratio in MRIMagnetic Resonance in Medicine, 1998
- Simulation ofB1Field Distribution and Intrinsic Signal-to-Noise in Cardiac MRI as a Function of Static Magnetic FieldJournal of Magnetic Resonance, 1997
- The intrinsic signal‐to‐noise ratio in NMR imagingMagnetic Resonance in Medicine, 1986