Metabolic Control With the Bio-inspired Artificial Pancreas in Adults With Type 1 Diabetes
Open Access
- 17 November 2015
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 10 (2) , 405-413
Abstract
Background: The Bio-inspired Artificial Pancreas (BiAP) is a closed-loop insulin delivery system based on a mathematical model of beta-cell physiology and implemented in a microchip within a low-powered handheld device. We aimed to evaluate the safety and efficacy of the BiAP over 24 hours, followed by a substudy assessing the safety of the algorithm without and with partial meal announcement. Changes in lactate and 3-hydroxybutyrate concentrations were investigated for the first time during closed-loop. Methods: This is a prospective randomized controlled open-label crossover study. Participants were randomly assigned to attend either a 24-hour closed-loop visit connected to the BiAP system or a 24-hour open-loop visit (standard insulin pump therapy). The primary outcome was percentage time spent in target range (3.9-10 mmol/l) measured by sensor glucose. Secondary outcomes included percentage time in hypoglycemia (10 mmol/l). Participants were invited to attend for an additional visit to assess the BiAP without and with partial meal announcements. Results: A total of 12 adults with type 1 diabetes completed the study (58% female, mean [SD] age 45 [10] years, BMI 25 [4] kg/m2, duration of diabetes 22 [12] years and HbA1c 7.4 [0.7]% [58 (8) mmol/mol]). The median (IQR) percentage time in target did not differ between closed-loop and open-loop (71% vs 66.9%, P = .9). Closed-loop reduced time spent in hypoglycemia from 17.9% to 3.0% ( P < .01), but increased time was spent in hyperglycemia (10% vs 28.9%, P = .01). The percentage time in target was higher when all meals were announced during closed-loop compared to no or partial meal announcement (65.7% [53.6-80.5] vs 45.5% [38.2-68.3], P = .12). Conclusions: The BiAP is safe and achieved equivalent time in target as measured by sensor glucose, with improvement in hypoglycemia, when compared to standard pump therapy.Keywords
This publication has 29 references indexed in Scilit:
- The pharmacokinetics and pharmacodynamics of rapid‐acting insulin analogues and their clinical consequencesDiabetes, Obesity and Metabolism, 2012
- Normal Reference Range for Mean Tissue Glucose and Glycemic Variability Derived from Continuous Glucose Monitoring for Subjects Without Diabetes in Different Ethnic GroupsDiabetes Technology & Therapeutics, 2011
- Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studiesBMJ, 2011
- Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 DiabetesNew England Journal of Medicine, 2010
- Novel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 DiabetesDiabetes Care, 2010
- Cellular modeling: insight into oral minimal models of insulin secretionAmerican Journal of Physiology-Endocrinology and Metabolism, 2010
- MD-Logic Artificial Pancreas SystemDiabetes Care, 2010
- Sustained Benefit of Continuous Glucose Monitoring on A1C, Glucose Profiles, and Hypoglycemia in Adults With Type 1 DiabetesDiabetes Care, 2009
- Continuous Glucose Monitoring and Intensive Treatment of Type 1 DiabetesNew England Journal of Medicine, 2008
- Feasibility of Automating Insulin Delivery for the Treatment of Type 1 DiabetesDiabetes, 2006