Spot scanning proton therapy minimizes neutron dose in the setting of radiation therapy administered during pregnancy
Authors:
Open Access
- 8 September 2016
- journal article
- case report
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (5) , 366-376
Abstract
This is a real case study to minimize the neutron dose equivalent (H) to a fetus using spot scanning proton beams with favorable beam energies and angles. Minimum neutron dose exposure to the fetus was achieved with iterative planning under the guidance of neutron H measurement. Two highly conformal treatment plans, each with three spot scanning beams, were planned to treat a 25-year-old pregnant female with aggressive recurrent chordoma of the base of skull who elected not to proceed with termination. Each plan was scheduled for delivery every other day for robust target coverage. Neutron H to the fetus was measured using a REM500 neutron survey meter placed at the fetus position of a patient simulating phantom. 4.1 and 44.1 were measured for the two initial plans. A vertex beam with higher energy and the fetal position closer to its central axis was the cause for the plan that produced an order higher neutron H. Replacing the vertex beam with a lateral beam reduced neutron H to be comparable with the other plan. For a prescription of 70 Gy in 35 fractions, the total neutron H to the fetus was estimated to be 0.35 mSv based on final measurement in single fraction. In comparison, the passive scattering proton plan and photon plan had an estimation of 26 and 70 mSv, respectively, for this case. While radiation therapy in pregnant patients should be avoided if at all possible, our work demonstrated spot scanning beam limited the total neutron H to the fetus an order lower than the suggested 5 mSv regulation threshold. It is far superior than passive scattering beam and careful beam selection with lower energy and keeping fetus further away from beam axis are essential in minimizing the fetus neutron exposure. PACS number(s): 87.53.Bn, 87.55.D-, 87.55.N-
Keywords
This publication has 28 references indexed in Scilit:
- Review of Radiation Therapy in the Pregnant Cancer PatientClinical Obstetrics and Gynecology, 2011
- Measurement of Neutron Dose Equivalent and its Dependence on Beam Configuration for a Passive Scattering Proton Delivery SystemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Neutron equivalent doses and associated lifetime cancer incidence risks for head & neck and spinal proton therapyPhysics in Medicine & Biology, 2009
- Assessment of organ-specific neutron equivalent doses in proton therapy using computational whole-body age-dependent voxel phantomsPhysics in Medicine & Biology, 2008
- Monte Carlo simulations of neutron spectral fluence, radiation weighting factor and ambient dose equivalent for a passively scattered proton therapy unitPhysics in Medicine & Biology, 2007
- Monte Carlo study of neutron dose equivalent during passive scattering proton therapyPhysics in Medicine & Biology, 2007
- Intensity-modulated radiation therapy, protons, and the risk of second cancersInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Simulation of organ-specific patient effective dose due to secondary neutrons in proton radiation treatmentPhysics in Medicine & Biology, 2005
- Evaluation of a Microdosimetric-based Neutron InstrumentRadiation Protection Dosimetry, 2000
- Secondary Dose Exposures During 200 MeV Proton TherapyRadiation Protection Dosimetry, 1997