The accuracy and safety aspects of a novel robotic needle guide manipulator to perform transrectal prostate biopsies
Authors:
- Martijn G. SchoutenMSMartijn G. Schouten
- Janneke AnsemsJAJanneke Ansems
- W. Klaas Jan RenemaWRW. Klaas Jan Renema
- Dennis BosboomDBDennis Bosboom
- Tom W. J. ScheenenTSTom W. J. Scheenen
- Jurgen J. FüttererJFJurgen J. Fütterer
- 17 August 2010
- journal article
- Published by Wiley in Medical Physics
- Vol. 37 (9) , 4744-4750
Abstract
Purpose: To introduce a new in‐house developed pneumatically controlled magnetic field compatible manipulator as an aid to perform magnetic resonance (MR)‐guided biopsies of the prostate.Methods: A pneumatic controlled manipulator with five degrees of freedom constructed of plastic to achieve magnetic field compatibility was developed to guide biopsies. A risk analysis, mechanical tests, and RF safety tests with respect to needle tip heating were performed to assure future patient safety and to meet standard safety requirements for the use in a medical environment. The accuracy of needle positioning with the needle guide manipulator to sample a predefined target was measured in agar phantoms on a 3 T whole body MR system. The in‐plane error was used to evaluate the accuracy, which is defined as the orthogonal distance between target and biopsy needle. The time for each step in the biopsy procedure was recorded to evaluate the procedure time. The influence of the insertion angle with respect to the static field of the MR scanner on the needle artifact was investigated.Results: The risk analyses met patient safety requirements. No RF induced local heating around the needle tip was observed. The average in‐plane error in 19 measurements was 3.0 mm (range 0–5.6 mm). The average time needed for manipulation to place the needle guide in the desired position was 5 min (range 3–8 min). Total procedure time was 30 min. The needle artifact size increases with the insertion angle with respect to the static field of the MR scanner.Conclusions: The new MR compatible manipulator can be used safely for patient care. It showed a high accuracy and short total procedure time, demonstrating great potential to improve the transrectal prostate biopsy procedure.Keywords
This publication has 28 references indexed in Scilit:
- Feasibility of 3T Dynamic Contrast-Enhanced Magnetic Resonance-Guided Biopsy in Localizing Local Recurrence of Prostate Cancer After External Beam Radiation TherapyInvestigative Radiology, 2010
- MRI-guided robotic system for transperineal prostate interventions: proof of principlePhysics in Medicine & Biology, 2010
- Thirty-Two-Channel Coil 3T Magnetic Resonance-Guided Biopsies of Prostate Tumor Suspicious Regions Identified on Multimodality 3T Magnetic Resonance Imaging: Technique and FeasibilityInvestigative Radiology, 2008
- The case for MR‐compatible robotics: a review of the state of the artThe International Journal of Medical Robotics and Computer Assisted Surgery, 2008
- Transperineal Prostate Intervention: Robot for Fully Automated MR Imaging—System Description and Proof of Principle in a Canine ModelRadiology, 2008
- Cancer Statistics, 2008CA: A Cancer Journal for Clinicians, 2008
- “MRI Stealth” robot for prostate interventionsMinimally Invasive Therapy & Allied Technologies, 2007
- MRI-Guided Biopsy of the Prostate Increases Diagnostic Performance in Men with Elevated or Increasing PSA Levels after Previous Negative TRUS BiopsiesEuropean Urology, 2006
- Interventional robotic systems: Applications and technology state‐of‐the‐artMinimally Invasive Therapy & Allied Technologies, 2006
- Prostate cancer detection and tumor characteristics in men with multiple biopsy sessionsProstate Cancer and Prostatic Diseases, 2004