Portable two-filter dual-flow-loop 222Rn detector: stand-alone monitor and calibration transfer device
Open Access
- 18 May 2022
- journal article
- research article
- Published by Copernicus GmbH in Advances in Geosciences
- Vol. 57, 63-80
Abstract
Little overlap exists in the required capabilities of 222Rn (radon) monitors for public health and atmospheric research. The former requires robust, compact, easily transportable instruments to characterise daily to yearly variability >100 Bq m−3, whereas the latter requires static instruments capable of characterising sub-hourly variability between 0.1 and 100 Bq m−3. Consequently, detector development has evolved independently for the two research communities, and while many radon measurements are being made world-wide, the full potential of this measurement network can't be realised because not all results are comparable. Development of a monitor that satisfies the primary needs of both measurement communities, including a calibration traceable to the International System of Units (SI), would constitute an important step toward (i) increasing the availability of radon measurements to both research communities, and (ii) providing a means to harmonize and compare radon measurements across the existing eclectic global network of radon detectors. To this end, we describe a prototype detector built by the Australian Nuclear Science and Technology Organisation (ANSTO), in collaboration with the EMPIR 19ENV01 traceRadon Project and Physikalisch-Technische Bundesanstalt (PTB). This two-filter dual-flow-loop radon monitor can be transported in a standard vehicle, fits in a 19′′ instrument rack, has a 30 min temporal resolution, and a detection limit of ∼0.14 Bq m−3. It is capable of continuous, long-term, low-maintenance, low-power, indoor or outdoor monitoring with a high sensitivity and an uncertainty of ∼15 % at 1 Bq m−3. Furthermore, we demonstrate the successful transfer of an SI traceable calibration from this portable monitor to a 1500 L two-filter radon monitor under field conditions.Keywords
Funding Information
- European Metrology Programme for Innovation and Research (19ENV01 traceRadon)
This publication has 56 references indexed in Scilit:
- Constraining annual and seasonal radon-222 flux density from the Southern Ocean using radon-222 concentrations in the boundary layer at Cape GrimTellus B: Chemical and Physical Meteorology, 2013
- Evaluation of anthropogenic emissions of carbon monoxide in East Asia derived from the observations of atmospheric radon-222 over the western North PacificAtmospheric Chemistry and Physics, 2012
- Fraction of the positive 218Po and 214Pb clusters in indoor airRadiation Protection Dosimetry, 2005
- Observations of atmospheric variability and soil exhalation rate of radon-222 at a Russian forest site. Technical approach and deployment for boundary layer studiesTellus B: Chemical and Physical Meteorology, 2002
- Verification of German methane emission inventories and their recent changes based on atmospheric observationsJournal of Geophysical Research: Atmospheres, 1999
- Baseline radon detectors for shipboard use: Development and deployment in the First Aerosol Characterization Experiment (ACE 1)Journal of Geophysical Research: Atmospheres, 1998
- Radon-222 time series measurements in the Antarctic peninsula (1986-1987)Tellus B: Chemical and Physical Meteorology, 1990
- A High-sensitivity Rn Detector Incorporating A Particle GeneratorHealth Physics, 1985
- Two-filter monitor for atmospheric 222RnReview of Scientific Instruments, 1980
- The Variation with Meteorological Conditions of the Amount of Radium Emanation in the Atmosphere, in the Soil Gas, and in the air Exhaled from the Surface of the Ground, at ManilaPhysical Review B, 1915