Investigation of optimum anti-scatter grid selection for digital radiography: physical imaging properties and detectability of low-contrast signals
Authors:
- 8 August 2012
- journal article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 6 (1) , 54-60
Abstract
Our aim in this study was to evaluate the effect of the grid variations on the imaging performance for a computed radiographic system under identical exposure condition. Digital radiographies using a 20-cm Lucite phantom were performed without grid and with grid ratios of 5:1, 8:1, 10:1, 12:1, and 14:1. The scatter fraction, the incident dose to the image receptor, the Wiener spectrum (WS), and the noise-equivalent quanta (NEQ) were measured. Visibility of low-contrast signals was evaluated using a contrast-detail phantom. The scatter fractions decreased considerably with an increase in the grid ratio. On the other hand, the WSs were increased (the noise property deteriorated) as the grid ratio increased due to a decreased incident dose to the image receptor under the identical exposure condition. The NEQs were improved as the grid ratio increased. The high grid ratios provided higher low-contrast detectability compared to the low grid ratios. Our results indicated that the removal of scattered radiation was very effective in improvement of the NEQ in the digital system under the identical exposure condition.Keywords
This publication has 40 references indexed in Scilit:
- Noise Power SpectrumJapanese Journal of Radiological Technology, 2009
- Digital radiography: The balance between image quality and required radiation doseEuropean Journal of Radiology, 2009
- Proposal and Verification of Presampled MTF Measurement by the Simple Analysis Method Using the Edge MethodJapanese Journal of Radiological Technology, 2008
- Imaging properties of digital magnification radiographyMedical Physics, 2006
- Generalizing the MTF and DQE to include x-ray scatter and focal spot unsharpness: Application to a new microangiographic systemMedical Physics, 2005
- Comparison of an amorphous silicon/cesium iodide flat‐panel digital chest radiography system with screen/film and computed radiography systems — A contrast‐detail phantom studyMedical Physics, 2001
- Clinical evaluation of a new stationary high strip density antiscatter grid in comparison with a conventional moving grid: influence on image quality and patient radiation doseEuropean Radiology, 2001
- Physical and clinical evaluation of a 2,048 × 2,048-matrix image intensifier TV digital imaging system in bone radiographyAcademic Radiology, 1996
- Basic imaging properties of a computed radiographic system with photostimulable phosphorsMedical Physics, 1989
- Investigation of the performance of antiscatter grids: Monte Carlo simulation studiesPhysics in Medicine & Biology, 1982