Drug-Eluting Stent Design is a Determinant of Drug Concentration at the Endothelial Cell Surface
Authors:
- 28 January 2016
- journal article
- research article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 44 (2) , 302-314
Abstract
Although drug-eluting stents (DES) have greatly reduced arterial restenosis, there are persistent concerns about stent thrombosis. DES thrombosis is attributable to retarded vascular re-endothelialization due to both stent-induced flow disturbance and the inhibition by the eluted drug of endothelial cell proliferation and migration. The present computational study aims to determine the effect of DES design on both stent-induced flow disturbance and the concentration of eluted drug at the arterial luminal surface. To this end, we consider three closed-cell stent designs that resemble certain commercial stents as well as three “idealized” stents that provide insight into the impact of specific characteristics of stent design. To objectively compare the different stents, we introduce the Stent Penalty Index (SPI), a dimensionless quantity whose value increases with both the extent of flow disturbance and luminal drug concentration. Our results show that among the three closed-cell designs studied, wide cell designs lead to lower SPI and are thus expected to have a less adverse effect on vascular re-endothelialization. For the idealized stent designs, a spiral stent provides favorable SPI values, whereas an intertwined ring stent leads to an elevated SPI. The present findings shed light onto the effect of stent design on the concentration of the eluted drug at the arterial luminal surface, an important consideration in the assessment of DES performance.Keywords
Funding Information
- National Research Foundation of Korea (KR)
- AXA Research Fund (FR)
This publication has 44 references indexed in Scilit:
- Modeling the transport of drugs eluted from stents: physical phenomena driving drug distribution in the arterial wallBiomechanics and Modeling in Mechanobiology, 2014
- Comparison of Near-Wall Hemodynamic Parameters in Stented Artery ModelsJournal of Biomechanical Engineering, 2009
- Arterial geometry, flow pattern, wall shear and mass transport: potential physiological significanceJournal of The Royal Society Interface, 2008
- Effects of different stent designs on local hemodynamics in stented arteriesJournal of Biomechanics, 2008
- Vascular Responses to Drug Eluting StentsArteriosclerosis, Thrombosis, and Vascular Biology, 2007
- Treatment of In-Stent Restenosis — Back to the Future?New England Journal of Medicine, 2006
- Experimental and Computational Flow Evaluation of Coronary StentsAnnals of Biomedical Engineering, 2000
- Vascular endothelium responds to fluid shear stress gradients.Arteriosclerosis and Thrombosis: A Journal of Vascular Biology, 1992
- Flow patterns and spatial distribution of atherosclerotic lesions in human coronary arteries.Circulation Research, 1990
- Turbulent fluid shear stress induces vascular endothelial cell turnover in vitro.Proceedings of the National Academy of Sciences, 1986